![GHM GREISINGER G 1111 Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/ghm/greisinger-g-1111/greisinger-g-1111_operating-manual_2220431014.webp)
5 | Operation
G 1111
14 / 28
B-H88.0.11.DB2-1.0
Representation
Call up menu
Hold
1s
Release
Release
_
Outcome of an action
The
adjustment menu
is closed after the last parameter.
5.2.5 Parameters of the adjustment menu
_
Description
Zero point
_
– For an optimal zero point calibration, a vacuum pump and a precise reference
device are needed.
– The zero point correction is used together with the gradient correction primarily for
compensation of sensor deviations. The entry takes place in the display unit.
Gradient
– A pressure reference is necessary for the gradient adjustment, e.g. in the form of a
pressure source and a reference display device.
– The gradient correction is used together with the zero point correction primarily for
compensation of sensor deviations.
_
Prerequisite
– The
adjustment menu
_
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the
Up button
or
Down button
.
3. The available configuration options are listed for each parameter in the following
representation.
_
Representation
Parameter
Values
Meaning
_
Zero point correction
PR
.
OF
0
.
0
No offset
PR
.
OF
-5
.
0
..
5
.
0
Offset active
_
Gradient compensation specification
PR
.
SL
0
No slope
PR
.
SL
-
5
.
00
..
5
.
00
Slope active
_
Formula
Sensor zero point correction offset
PR
.
OF
A zero point shift can be made for the measurement.
Displayed value = measured value – offset
Standard setting: 0.0, i.e. no correction is made.
Sensor gradient correction
PR
.
SL
The gradient of the measurement can be influenced with this factor (factor is in %):
Displayed value = measured value * (1+Pr.SL/100)
Standard setting: 0.000, i.e. no correction is made.
_
Outcome of an action
The changed value is saved and the
adjustment menu
is closed.
STOR
appears in the
display.