Operation | 7

B-H87.0.01.DB200-2.0

17 / 24

4. When you have selected the desired parameter, change the parameter to the de-

sired value with the 

Up button

 and the 

Down button

.

5. The changes are saved after running through the entire 

Configuration

 menu. 

STOR

appears in the display. The 

Configuration

 menu can be exited from any arbitrary

parameter by pressing and holding the 

Function key

 for 2 seconds. The changes

made up that point are saved.

_

Representation

Call up menu

Next parameter Change value

Save changes

Discard
changes

 

2s

Press: Single
step

Hold: Fast
change

2s

2s

Product is
switched off

_

Outcome of an action

The 

Configuration

 menu is closed after the last parameter.

 NOTE

If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.

7.2.3 Configuring parameters of the configuration menu

_

Description

_

The following representation shows the available parameters and various configura-
tion options.

_

Prerequisite

– The 

Configuration

 menu is open. See 

Opening the configuration menu [

}

 p. 16]

.

_

Instruction

1. Select the desired parameter you would like to configure.

2. Adjust the desired configuration in the selected parameter with the 

Up button

 or

Down button

.

3. The available configuration options are listed for each parameter in the following

representation.

_
_

Representation

Parameter

Values

Meaning

 

_

Input

INP

(OND

Measured variable - conductivity

SAL

Measured variable - salt content / salinity

TDS

Measured variable - total dissolved solids

_

Faktor für TDS

CTDS

0

.

40

 .. 

1

.

00

Conversion factor for TDS measurement

_

Содержание G 1410

Страница 1: ...Operating manual G 1410 Handheld conductivity measuring device EN...

Страница 2: ...2 General information about conductivity measuring 11 5 3 Conductivity measurement 11 5 4 Total dissolved solids TDS measurement 12 5 5 Salt content salinity measurement 12 5 6 Electrodes measuring ce...

Страница 3: ...Table of contents B H87 0 01 DB200 2 0 3 24 8 Error and system messages 22 9 Disposal 23 10 Technical data 24...

Страница 4: ...is document disregarding safety notices assignment of inadequately qualified technical personnel and arbitrary modifications of the product Only carry out the maintenance and service tasks on this pro...

Страница 5: ...Formula Some instructions include a formula for a general understanding of a configuration programming or a setting of the product Outcome of an action Result consequence or effect of an instruction...

Страница 6: ...he device specific safety instructions for this docu ment are observed If these notices are disregarded personal injury or death as well as property damage can occur _ DANGER Incorrect area of applica...

Страница 7: ...have ad equate knowledge of the measuring process and the significance of the measure ments This document makes a valuable contribution to this The instructions in this document must be understood obs...

Страница 8: ...speed and reliability in a compact ergonomic housing Additional impressive features include the dust proof and waterproof design in accord ance with IP 65 67 and the 3 line illuminated display which o...

Страница 9: ...in max hold _ Auxiliary display Corresponding temperature value for the value shown in the main display If applicable alternating with the temperature compensation 4 3 Operating elements _ On Off butt...

Страница 10: ...e G 1410 10 24 B H87 0 01 DB200 2 0 Function key Press briefly Freeze measurement Hold Return to measurement display Call up next parameter Long press 2s Open menu frozen measurement is displayed Clos...

Страница 11: ...stored dry or in water After dry storage wetting time will be prolonged slightly If changing over from one liquid to another with conductivit ies varying widely make sure to properly rinse and shake d...

Страница 12: ...calibration to known comparison solutions or with actual evaporation of a certain amount of liquid with measured conductivity and subsequent weight of the total dissolved solids 5 5 Salt content salin...

Страница 13: ...testing and calibration solutions such as GKL 100 101 and 102 are available for this purpose In normal application conditions semi annual testing is re commended Adjustment of the measuring input p 19...

Страница 14: ...arrow range This is only necessary for normal use See Adjust ment of the measuring input p 19 6 2 Battery 6 2 1 Battery indicator _ If the empty frame in the battery display blinks the batteries are d...

Страница 15: ..._ Prerequisites The product is switched off _ Instruction A B C D Fig 3 Batteriewechsel 1 Unscrews the Phillips screws A and remove the cover 2 Carefully replace the two Mignon AA batteries B Ensure...

Страница 16: ...duct is now ready for measurement NOTE The device is calibrated at the factory and ready for measurement The offset and gradient correction of the temperature measurement and gradient correction of th...

Страница 17: ...ct is switched off without saving the configuration the last save value is re produced on the next start up of the product 7 2 3 Configuring parameters of the configuration menu _ Description _ The fo...

Страница 18: ...TE OFF Backlight deactivated 15 30 60 120 240 Automatic shut off of the backlight after a selected time in seconds during which no buttons have been pressed ON No automatic shut off of the backlight _...

Страница 19: ...the Configuration menu Release the Down button The display shows the first parameter 3 By briefly pressing the Function key you can scroll through the parameters Select the parameter you would like to...

Страница 20: ...C Display measured value T OF 1 T SL 100 _ Gradient correction F Display measured value 32 F T OF 1 T SL 100 32 F _ Gradient correction Display measured value T SL _ Example calculation _ Temperature...

Страница 21: ...e achieved when the control solution temperature is ad justed to 25 C With different temperatures the table value of the corresponding tem perature must be used as a setpoint _ Outcome of an action Th...

Страница 22: ...ur ing range Send in for repair Send in for repair No display unclear char acters or no response when but tons are pressed Battery depleted System error Product is defective Battery depleted Error in...

Страница 23: ...ectives applicable at the time must be observed _ NOTE The device must not be disposed of with household waste Return it to us freight pre paid We will then arrange for the proper and environmentally...

Страница 24: ...ight 108 54 28 mm without measuring cell or kink protection 180 g incl battery and measuring cell Operating conditions 20 to 50 C 0 to 95 r h temporarily 100 r h Storage temperature 20 to 70 C Power s...

Отзывы: