Operation | 7
B-H87.0.01.DB200-2.0
17 / 24
4. When you have selected the desired parameter, change the parameter to the de-
sired value with the
Up button
and the
Down button
.
5. The changes are saved after running through the entire
Configuration
menu.
STOR
appears in the display. The
Configuration
menu can be exited from any arbitrary
parameter by pressing and holding the
Function key
for 2 seconds. The changes
made up that point are saved.
_
Representation
Call up menu
Next parameter Change value
Save changes
Discard
changes
2s
Press: Single
step
Hold: Fast
change
2s
2s
Product is
switched off
_
Outcome of an action
The
Configuration
menu is closed after the last parameter.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
7.2.3 Configuring parameters of the configuration menu
_
Description
_
The following representation shows the available parameters and various configura-
tion options.
_
Prerequisite
– The
Configuration
Opening the configuration menu [
.
_
Instruction
1. Select the desired parameter you would like to configure.
2. Adjust the desired configuration in the selected parameter with the
Up button
or
Down button
.
3. The available configuration options are listed for each parameter in the following
representation.
_
_
Representation
Parameter
Values
Meaning
_
Input
INP
(OND
Measured variable - conductivity
SAL
Measured variable - salt content / salinity
TDS
Measured variable - total dissolved solids
_
Faktor für TDS
CTDS
0
.
40
..
1
.
00
Conversion factor for TDS measurement
_