Connecting the pH-Electrode
Connect the sensor connection cable into the corresponding BNC connector socket.
Before using for KH measurement, the pH electrode must
first
be calibrated. Please note the
following instructions in this manual.
4.11
Calibrating the pH-Electrode
Calibration is always necessary with new sensors. Also, the calibration should be repeated
from time to time (approximately every 2-4 weeks) due to the aging process of the sensor.
Please also observe the information enclosed with the sensor.
Connect the pH-Sensor.
If you have not already done so, check the check box for
KH Director enabled
and save the
changes.
DANGER
•
Sensor plugs must not be wet or damp when connecting to the
socket.
•
Do not use force.
•
Products that are already powered should never be pulled by the
cable. This may cause malfunction or damage the connected
products and the
KH Director
.
•
The KH Director may NEVER be used without a built-in pH
electrode!
•
A repair caused by this is not a guarantee and is therefore subject
to a repair charge.
DANGER
•
Please allow some time for the calibration
•
Please follow the instructions of the
GHL Control Center
carefully
•
Before immersing the sensor in a calibration solution, the sensor
must be carefully dried with a dry, lint-free paper towel.