The Signet 3719 pH/ORP Wet-Tap allows installation and removal of pH or 
ORP electrodes without the need for process shutdown during routine electrode 
maintenance and calibration. Process isolation is achieved with a double O-ring 
seal on a unique and compact retraction assembly; no separate valve is required. 
A cam-activated automatic locking mechanism, SafeLoc™, and the short stroke 
design help to assure operator safety.

English

WARNING!

If used in conditions that exceed recommended 
pressure ratings, this product can pose a serious 
hazard.
We urge customers to read the specifi cations 
carefully before installing and operating this product.
Improper use can cause components and process 
liquids to be expelled at high speeds and cause 
serious personal injury.

*3-3719.090*

3-3719.090    Rev. 18   06/19

Signet 3719 pH/ORP Wet-Tap

Description

Table of Contents

Operating Instructions

• 

English

• 

Deutsch

• 

Français

• 

Español

Warranty Information ......................................................... 2
Product Registration .......................................................... 2
Safety Information ............................................................. 2
Dimensions ........................................................................ 2
Specifi cations .................................................................... 3
System Overview............................................................... 3
Installation ......................................................................3-5

Location, Orientation and Required Clearance

 ............3-4

 

For pipe sizes 2.5 to 12 in. .......................................... 4

 

For pipe sizes less than 2.5 in. .................................... 4

 

Wet-Tap into Fitting...................................................... 5

 

Electrode Installation ................................................... 5

Electrode Removal ............................................................ 6
Electrode Maintenance ...................................................... 6
Troubleshooting ................................................................. 7
Replacing O-rings .............................................................. 8
Ordering Information........................................................ 12

Содержание Signet 3719

Страница 1: ...omers to read the speci cations carefully before installing and operating this product Improper use can cause components and process liquids to be expelled at high speeds and cause serious personal in...

Страница 2: ...s or application failures e g high temperature chemical poisoning dry out or mishandling e g broken glass damaged membrane freezing and or extreme temperatures Thank you for purchasing the Signet line...

Страница 3: ...ee ordering information on page 8 DryLoc refers to the electrode connector style d 3 2751 1 or 2760 DryLoc pH ORP Sensor Electronics Preampli er e 3 2751 2 electronics NOTE Output signal options See s...

Страница 4: ...g a ow cell with standard piping components One simple solution using a tee tting and reducer bushings is shown in the example below Many similar con gurations are conceivable in a wide variety of mat...

Страница 5: ...peration Use an appropriate thread sealant to prevent leaks The piping system can now be safely pressurized Inspect the installation for leaks 3 4 Electrode Piston Electrode Installation Before instal...

Страница 6: ...ean any scaling and debris found on or in the piston tip and surrounding area 2 3 4 5 1 5 3 4 Turn the locking shroud turn counterclockwise and lift up to remove it from the Wet Tap assembly 5 Remove...

Страница 7: ...the tip of the piston and into each O ring groove 6 Lubricant 7 countersink hole Stainless Steel set screw hole Locate the countersink hole on the nosepiece Verify that it is clean and free of debris...

Страница 8: ...PT 3 3719 12 159 000 806 pH ORP Wet Tap ISO 7 1 R 1 5 3 3719 22 159 000 807 Wet Tap Assembly ISO 7 1 R 2 Parts and Accessories 3 2756 WTP 159 001 390 Electrode pH DryLoc plastic bulb Pt1000 Wet Tap 3...

Отзывы: