19
John Bunn O Premier Oxygen Concentrator
8 sPeCIfICatIOns
• Flow Range: 1 to 5L/min
• Oxygen Concentration: 1-4 LPM 94% ± 2%
5 LPM 92% ± 2%
• Maximum Outlet Pressure: 6.52 psi ± .065 psi
(45kPa ± 4.5kPa)
• Pressure Relief Mechanism Operational at:
36.26 psi ± 3.63 psi (250kPa ± 25kPa)
• Altitude: Up to 5,997 feet (1,828 meters) above sea level
without degradation of concentration levels. Efficiency
of 90% from 5,997 to 13,123 feet (1,828 to 4,000 meters)
above sea level.
• Sound Level: ≤ 55dB
• Power Supply: JB0160-010/JB0160-015: AC120V±10%
(60Hz) (4.6amps). JB0160-010-220/JB0160-015-220:
AC220V±10% (50Hz) (2.2amps)
• Input Power: 400 Watts Average
• Weight: 57lbs (26kg)
• Dimensions: H 27” x W 18” x D 15”
• Minimum Warm-Up Time Before Use: 30 minutes
• Electrical Classification: II Class B Type
• System: Continuous
• JB0160-015/JB0160-015-220 only: O2 sensor (electro-
chemical device): Yellow fault light illuminates when O
2
concentration is greater than 71% and less than 82%;
system works. Red warning light illuminates when O
2
concentration is less than 71%; system shuts off.
normal Operation Conditions
• Temperature range: 41
°
F to 104
°
F (5
°
C to 40
°
C)
• Relative humidity: 20% to 60%
• Atmospheric pressure: 12.47 psi to 15.37 psi
(86kPa to 106kPa)
Содержание John Bunn O2 Premier JB0160-010
Страница 23: ...23 John Bunn O2 Premier Oxygen Concentrator Notes...
Страница 26: ...Affix first class postage here GF Health Products Inc 2935 Northeast Parkway Atlanta GA 30360...
Страница 27: ......