5 MaIntenanCe
Cleaning the Cabinet
Clean the cabinet at least once a month.
1. Turn off the power switch.
2. Clean the cabinet with a mild household cleaner and a
non-abrasive cloth or sponge. Do not spill liquid inside
the cabinet.
Cleaning or Replacing the filters (3 types)
Clean or replace the filters as often as specified in the
following paragraphs in order to protect the compressor
and extend the concentrator’s life. Replacement filters are
available from your Graham-Field equipment provider.
Bottom filter
The bottom filter does not typically require cleaning, but
it may if operated in extremely dirty conditions. Perform a
visual inspection of the bottom filter every 2 weeks; if the
filter is dirty, clean it. The bottom filter is located at the
bottom rear of the cabinet (see Figure 1). To clean:
1. Remove the bottom filter
from the squares carefully
with your fingers.
2. Inspect the filter; if any part
of it is cracked or broken,
replace it instead of clean-
ing it.
3. If cleaning, rinse filter well
with clear water until all dirt is removed.
4. Dry filter by squeezing between two absorbent towels.
Wait for filter to dry.
When dry, install filter carefully as
shown in Figure 1.
s
Damage: Do not operate the concentrator without the
filters installed, or while filters are wet. These actions
could permanently damage the concentrator.