Optional Equipment
7–2
Temperature Guarantee
DESCRIPTION
All phases of the wash cycle except Soak and Exhaust have the option to
select a Temperature Guarantee Mode. When selected, this causes the unit
to fill and spray water normally for that phase, without allowing the main
timer to start counting down until the setpoint temperature has been
attained. If at any time the sump temperature should fall below the set point,
the main timer stops until the temperature exceeds the setpoint. Failure to
do so in a set period of time generates a “TOO LONG TO HEAT” alarm.
USER CONTROL
Program the desired temperature into the user program and turn on the
guarantee in the desired phase(s).
Controls
TOUCH CONTROL PANEL (TCP)
A solid state microcomputer control system monitors and automatically
controls all process operations and functions. Cycle phase times and
temperatures, and other key process parameters are programmable and
may be locked in by supervision. The twelve-cycle menu of treatment
processes may be programmed and retained to permit operating personnel
to accommodate a wide variety of load and processing requirements. Cycle
programming may be controlled by a supervisory access code to ensure
process integrity. Each cycle program may be reviewed and printed on
demand.
The optional Strip Chart Printer must be selected to have this feature.
Programming is in standard military time, Fahrenheit or Centigrade
temperature. Times and temperature are expressed in seconds and one
degree increments respectively. An internal battery backs up all cycle
memory.
A highly visible color touch panel screen displays cycle program data on
demand and real time in process cycle performance. All cycle deviations are
alarmed both visually and audibly, recorded and must be acknowledge by
the user.
A built-in service diagnostic program, accessible by service access code, is
included and displayed to permit system calibration and verification of
satisfactory component operation. Service access to the control panel is
from the front of the control terminal to permit further ease of maintenance.
LED CONTROL PANEL
A solid state microcomputer control system monitors and automatically
controls all process operations and functions. Cycle phase time and
temperatures, and other key process parameters are programmable and
can be locked in by supervision. The cycle menu of treatment processes
can be programmed and retained to permit operating personnel to
accommodate a wide variety of load and processing requirements. Cycle
programming can be controlled by a supervisory access code to insure
NOTE