Troubleshooting
5–2
Shutting Down the Washer
In an emergency, it may be necessary to remove electrical power supplied
to the washer or to turn off the steam and water supplies. All users should
know the location of the main circuit breaker to disconnect the main
electrical power and the steam and water supply shutoff valves at the
washer site.
The Electrical Disconnect Switch de-energizes the controls only; it does not
remove all power from the washer.
Pump pressure is lower than
normal (20-30psi).
1.
Spray jets missing.
2.
Header plugs are not in place.
3.
Pump seals are worn (check
for water under pump).
4.
Self cleaning debris filter may
be plugged.
5.
Pump connection to header is
leaking, allowing air into
pump.
1.
Replace spray jets as
necessary
2.
Replace as necessary.
3.
Replace as necessary.
4.
See “INSPECT THE
AUTOMATIC SELF-
CLEANING DEBRIS FILTER
SCREEN AND SUMP” on
page 4-3
5.
Replace hose or tighten
clamps.
Water is leaking from washer
1.
Door(s) not fully closed and
secured.
2.
Door gasket damaged.
3.
Window gasket damaged.
4.
Piping is leaking.
5.
Header is not installed.
6.
Duct work leaking, water
running down side of machine.
1.
Fully close and secure the
door(s).
2.
Replace the door gasket.
3.
Replace the window gasket.
4.
Check piping and valves.
Tighten or replace as
necessary.
5.
Install the header.
6.
Repair the duct work.
Cycles do not begin when the start
button is pressed.
1.
Water level is low.
2.
Machine not up to
temperature.
3.
Cycle parameters not set
properly.
1.
Inspect and clean lever control
probes.
2.
Check steam supply. Check
utility valves.
3.
Check program settings.
PROBLEM
PROBABLE CAUSE
CORRECTION
WARNING
Содержание Castle MTP 1800 Series
Страница 1: ...SERIES 1800 CAGE AND BOTTLE WASHER OPERATING MANUAL 61301604086 Rev A...
Страница 2: ......
Страница 68: ...Troubleshooting 5 6...
Страница 74: ...Installation Instructions 6 6...
Страница 92: ...Optional Equipment 7 18...
Страница 97: ......