Getinge 502606400 Скачать руководство пользователя страница 1

GETINGE 86-SERIES

INSTALLATION MANUAL
502606400

SEV0725001-

Содержание 502606400

Страница 1: ...GETINGE 86 SERIES INSTALLATION MANUAL 502606400 SEV0725001...

Страница 2: ...______________________________ 5 Attention symbols___________________________________________ 5 INSTALLATION_ __________________________________________6 Wall mounted model_ __________________________...

Страница 3: ...by Getinge Disinfection AB The customer is responsible for ensuring that an Installation Qualification an Operating Qualification and a Performance Qualification according to EN ISO 15883 are carried...

Страница 4: ...tions about what to do if the substance comes into contact with the eyes or skin or if vapors are inhaled storage of packs and sorting of empty packs for disposal The machine must be connected in acco...

Страница 5: ...e machine into service Incorrect use may result in damage and injury Product liability Any modification or incorrect use of the equipment without the approval of Getinge Dis infection AB invalidates G...

Страница 6: ...machine at the chosen location and adjust the feet so that the machine is stable Bottom frame V318 Figure 1 Adjustment Using a spirit level check as shown in Figure 1 that the machine is level within...

Страница 7: ...other they must be adjusted to the same height on installation Where a plinth plate is used the height from the floor to the underside of the ma chine must be at least 100 mm Set the machine horizonta...

Страница 8: ...supply to the machine should be fitted with a lockable isolator with a 3 mm break gap The isolator must be located in an easily accessible position on the wall It is important that the connection has...

Страница 9: ...PE 50 60Hz 10 4kW 32A FUSE 3x32A Steam heated 240 415V 3N PE 50Hz 11 4kW 18A FUSE 3x20A 240V 3 PE 60Hz 11 5kW 30A FUSE 3x30A electrically heated 230 400V 3N PE 50 60Hz 10 7kW 17A FUSE 3x20A 230V 3 PE...

Страница 10: ...e following requirements Where the pressure is lower than 100 kPa a separate feeder pump must be used At chamber depth 720 mm At chamber depth 800 mm Note Distilled or de ionized water must have a con...

Страница 11: ...400 mm above the floor 5 Steam connection 350 mm above the floor 6 Condensate connection 350 mm above the floor Chamber depth 720 mm Chamber depth 800 mm Clean side Soiled side Clean side Soiled side...

Страница 12: ...g the connections Connect the disinfector to a waste outlet with a capacity of at least 50 litres min Connect the waste as shown in Figure 5 Diameter of waste 50 mm Figure 6 Connecting a dryer Connect...

Страница 13: ...opened check that the clean side door is closed and locked If not proceed as follows Manually operated door Close the door and press W Automatically operated door Press W Place a wash rack in the wash...

Страница 14: ...rogram is complete as follows If the soiled side door cannot be opened check that the clean side door is closed Open and close the clean side door Press W Open the soiled side door Check that the clea...

Страница 15: ...4 3 9 kN m2 Floor area loading machine foot 347 kN m2 Connections Water consumption 8666 CM303 33 liters phase Note 1 8668 CM304 40 liters phase Note 1 Cold water Recommended water quality Drinking wa...

Страница 16: ...rature 160 C 320 F Consumption about 0 9 1 0 kg min at 300 kPa Steam condensate only for a machine with steam heated chamber Connection ISO G 1 2 Note 4 Waste outlet Dimension out 50 mm 2 Minimum capa...

Страница 17: ...Octave band frequency Hz 125 250 500 1000 2000 4000 8000 LwA LwAFmax Correction terms Kok 67 69 67 67 69 67 70 75 84 The calculated sound power level implies different sound levels LpA in different ty...

Страница 18: ...High concentrations of and high exposure to sanitary chemicals may cause corrosion and pitting on stainless steel Getinge Disinfection AB therefore recommends that water used in the pre rinsing wash i...

Страница 19: ...Page 19 of 20 Doc_INS Doc_502606400 Rel_A Lang_GB...

Страница 20: ...1 330 Italy Getinge S p A Via dei Buonvisi 61 D IT 00148 Roma info getinge it www getinge com it Phone 39 06656631 Netherlands Getinge B V Fruiteniersstraat 27 Zwijndrecht Postbus 1004 NL 3330 CA Zwij...