Electrical Requirements
Electrical Requirements are as follows:
LFGI-3USP 115V, 15 amp plug for one 10 ft. Power cord
LFGI-4USP 115V, 15 amp plug for one 10 ft. Power cord
LFGI-6USP 115V, 15 amp plug for one 10 ft. Power cord
It is essential that the equipment be properly grounded, it is a violation of your warranty to
operate without.
For service to the unit it is essential that the cord cap be easily accessible.
Optional Equipment affects Electrical Requirements in the following way-
Electrical Height Adjustment requires its own power receptacle and a 15 amp circuit on all
sized units.
Additional electrical components may be required based upon equipment used inside the
LFGI.
Certification
The Laminar Flow Glovebox /Isolator is a complete and self-contained unit whose
performance was tested and documented before shipping. It requires certification after the
unit has been installed and prior to operation. This is especially important after
transportation of the LFGI from our factory to your facility. The purpose of this retesting is
to verify that the HEPA filters, airflows and pressures are within limits and the unit is
performing properly.
It is always a good idea to review basic operation of your new LFGI with the certification
company personnel due to their knowledge of this type of equipment.
Certification may also be a Governmental and /or Professional requirement.
LFGI certification needs to be performed by a company that has the proper equipment and
training needed to test and measure HEPA filter performance. Certification of the LFGI
should be continued at least annually or as regulations dictate.
Please contact GERMFREE at 800-888-5357 for a list of Certification Companies in your
area if you do not already have this service.
16
Содержание LFGI-3USP
Страница 1: ...Laminar Flow LFGI CAI or CACI User Manual ...
Страница 33: ...33 ...
Страница 51: ...SECTION IV UNIT DIAGRAM Made in the USA and available for purchase on the GSA 51 ...
Страница 55: ...55 ...
Страница 59: ...59 ...
Страница 63: ...Fig 2 Air recycle plenum connection tip removed 63 ...
Страница 85: ... 863 248 2800 www sensocon com 20 Installation and Operation Manual Series A3 A4 ...