2.

 

Diffustik 

2.1

 

Description and Scope of Delivery 

2.1.1

 

Description 

 
The medical device Diffustik is a PC based system, which is used for the 
determination of the diffusion capacity in lung function measurements. 
It could be used as a single device or in combination with other 
products of Geratherm Respiratory. 
The Diffustik consists of a gas analyzer for carbon monoxide (CO) and 
methane (CH4), a demand valve with shutter unit as well as a housing 
with integrated electronic. The Diffustik connects to a PC via the 
integrated USB interface. The established Ergoflow Flow sensor 
principle is used for flow measurement.  
The Diffustik can only be used with the medical cart authorized by 
Geratherm Respiratory. The galvanic isolation of the electric power 
system is done by the medical cart. All peripheral equipment like 
computers, printers or monitors should only be connected to the 
electric power system of the cart. 
The Software Blue Cherry is delivered with the Diffustik. Its Standard 
version contains Measurements for SVC, FVC, CO-Diffusion as well as 
Pre/Post analysis and trend analysis. Furthermore it contains a powerful 
database for patient and test information and has a modular design, 
which offers the possibility of combination with other measurements. 
The modular and flexible software concept allows user specific 
configuration of Blue Cherry. A user may add to the predefined 
medication list, select between different predicted equations sets and 
add further predicted equations, define new parameters, configure 
results table and change patient data input view.  
Blue Cherry can be connected to office information software via a GDT 
interface. There are different export possibilities into Excel and XML. 
Using the standard Microsoft ®clipboard, all charts may be imported 
into standard software such as MS Word or MS PowerPoint. 
The protocol printing happens by the Windows Printer Interface. Every 
Windows compatible printer can be used. All test readings and 

Содержание Diffustik

Страница 1: ...Diffustik User Manual Version 1 2 2 Date 17th October 2014...

Страница 2: ...ON THE DIFFUSTIK 11 2 3 SECURITY 12 2 3 1 GENERAL SECURITY INFORMATION 12 2 3 1 1 NORMS AND GUIDELINES 12 2 3 1 2 ELECTROMAGNETIC COMPATIBILITY 13 2 3 2 SECURE USE OF DIFFUSTIK 13 2 3 3 CLEANING AND D...

Страница 3: ...be integrated into even the largest of requirements This document contains important information for the operation of Blue Cherry and Diffustik We strongly recommend reading this manual carefully in o...

Страница 4: ...ould only be connected to the electric power system of the cart The Software Blue Cherry is delivered with the Diffustik Its Standard version contains Measurements for SVC FVC CO Diffusion as well as...

Страница 5: ...ic unit with USB connector Gas analyzer for CO and CH4 Housing with integrated electronic Blue Cherry Software CD Modular diagnostic platform operating under Microsoft Windows environment Quick User G...

Страница 6: ...d disinfect the device The connectors of the Diffustik are located at the back side of the device Connect a grounded power supply cable to the AC 230V input 1 Now connect one of the bus connector 2 wi...

Страница 7: ...urces which are connected to a reliable protective earth conductor 3 pin cable grounded and which specifications meet the details described on the label on the rear of the housing Extension cords must...

Страница 8: ...t to the dovetail groove at the holder plate of the stand 3 After this plug the body of the electric unit 1 as far as it will go onto the closing head and connect the connection cable of the electric...

Страница 9: ...cable to the jack of the electric unit 2 2 3 Installation of the demand valve If the Bodystik includes the option CO Diffusion the demand valve has to be connected additionally Therefore move the dema...

Страница 10: ...connected accurately and without too much power The gas supply hose must only be connected to gas bottles with the following content up to 0 3 CH4 up to 0 3 CO up to 0 3 C2H2 optionally 21 O2 rest N2...

Страница 11: ...Diffustik Connect the provided power supply to the jack of the Diffustik Now connect the Diffustik via the provided RS232 cable to a free RS232 port at your computer 2 2 6 Power on the Diffustik First...

Страница 12: ...erstand this included user manual paying attention to the operating instructions Geratherm Respiratory is responsible for the security and reliability of function of the Diffustik as such only authori...

Страница 13: ...hones close to the device There may be an influence to medical devices The interference resistance is 3V m Do not use a higher level close to the Diffustik 2 3 2 Secure use of Diffustik The Diffustik...

Страница 14: ...Medical devices should not be used close to a patient always ensure a minimum distance of 1 5m between patient and non medical computer systems Leakage current may increase by connecting non medical...

Страница 15: ...on the device don t put any objects on top of the device Never move foreign materials into the housing Never place the device on uneven or unsecured surfaces The device should not be used in dirty env...

Страница 16: ...ensure all connections and cables are connected properly All additionally connected devices must comply with legal regulations and the total leakage current must not exceed the maximum allowed valued...

Страница 17: ...hin patient area and resource B in medical used room Resource A within patient area and resource B in medical used room See remarks below table Resource A within patient area and resource B in a non m...

Страница 18: ...ith local directives 2 3 6 Product labelling Diffustik Example of typical product labelling SN designates the serial number for the device The first 2 digits define the manufacturing year The example...

Страница 19: ...designates the serial number for the device The first 2 digits define the manufacturing year The example 13 shows the device was manufactured in 2013 The identifier 8 402 is a product specific code Th...

Страница 20: ...se Application component type BF according to DIN EN 60601 1 Indicates single use device Protection type II according to MDD Separate collection of electro and electronic devices IPX0 Indicates no pro...

Страница 21: ...intenance device Only minor maintenance is required to ensure trouble free operation Interval Maintenance work Once a day Flow sensor flow sensor housing and shutter unit cleaning and disinfect Once a...

Страница 22: ...be provided by Geratherm Respiratory or an authorized service partner of Geratherm Respiratory Geratherm Respiratory will not accept warranty claims on any unauthorised repairs 3 2 Warranty exclusion...

Страница 23: ...5 Hz 3 or 20ml s Volume Measurement range Accuracy 0 20L 3 or 50mL CO Analyser Measurement principle Measurement range Accuracy Resolution Infra red absorption 0 3000 ppm 1 24 bit CH4 Analyser Measure...

Страница 24: ...racy Resolution Infra red Absorption 0 3000 ppm 1 24 Bit CH4 Analyser Measurement principle Measurement range Accuracy Resolution Infrarot absorption 0 3000 ppm 1 24 bit Operating conditions Temperatu...

Отзывы: