Geratherm
®
Baby ColorChoice
®
in
Pink
or
Blue
Instructions for Use:
1. How to Switch between °F and °C
a) Press the On/Off button 2 times, one right after another.
b) Hold the second press down for more than 1.5 seconds.
c) Symbol °F or °C appears, with beep, on the display window.
2. How to Operate
a) Press down the On/Off button to start the thermometer. Then, a full display
(188.8
°
) ensures that the LCD window is in good function.
b) The next appearance ok signals normal operation after internal safe testing.
Otherwise, Err or --- alerts its malfunction.
c) Then, the ambient temperature or Lo
°F
( Lo
°C
) displays on window with °F or °C
blinking. The thermometer is now ready for a temperature taking. If there is no
temperature taking in the coming 3 minutes, the thermometer will be
automatically shut-off.
d) Put the probe head under the tongue, under the arm or in the rectum; according
to the methods provided (see 5. Methods of Use).
e) When the thermometer completes its temperature taking, °F or °C stops blinking
and the measured temperature displays on the LCD window with on series of
beeps.
f) The thermometer prolongs its process of temperature taking in the next 1 minute.
After that, the measured temperature is fixed and shown on LCD window until
auto shut-off in 9 minutes.
g) When the measured temperature is above or below the measuring range, the
LCD window displays Hi
°F
( Hi
°C
) or Lo
°F
( Lo
°C
).
3. How to Recall the Last Temperature
a) Keep pressing the On/Off button down approximately 1.5 seconds. The last
temperature will be recalled on the LCD window with “M” flag appeared at the
right-bottom corner. For example, 98.6
°F
( 37.0
°C
).
M
M
M
E