37
EN
Precautions
• Do not wrap the cuff around an arm to which medical
devices are attached, as this could lead to temporary
functional failure of the medical devices being used.
• Avoid any kinks in the air hose during use, as this
could lead to a continuous increase in the cuff pres-
sure and cause injury to the patient.
• This unit is not suitable for continuous monitoring dur-
ing medical emergencies or surgery. This unit cannot
be used simultaneously with high-frequency surgical
devices.
• This unit is not intended for the transportation of pa-
tients outside medical establishments.
• In order to avoid inaccurate measurements, do not
expose the blood pressure monitor to strong electro-
magnetic
fi
elds.
• The material of the cuff has been tested and meets the
requirements for the biological evaluation of medical
devices in accordance with the standards DIN EN ISO
10993-5 and DIN EN ISO 10993-10. The constituent
materials do not have the potential to produce any ir-
ritation or allergic reactions.
• Do not use the cuff on damaged skin.
• Do not use the device in the case of known allergies to
polyester or synthetic materials.
• In order to avoid accidental strangulation, keep the de-
vice away from children and do not lay the tube around
the neck.
• Do not connect the air hose to any other medical de-
vices, as this can lead to high pressure or conduct air
into intravascular systems, which may lead to injury.
• Before using the unit for the
fi
rst time, ensure that there
is no visible damage to the unit.
• Please operate the unit under the environmental condi-
tions described in the instructions. Otherwise, the per-
formance and service life of the blood pressure monitor
will be adversely affected.
• Please use only accessories that have been approved
by the manufacturer. Otherwise, damage may be
caused to the unit, the user may suffer injury or inac-
curate measurements may occur.
• Dispose of the unit, accessories and components in
accordance with local regulations.