10
Fault Indication and Repair
Simple errors which occur when using Spirostik can be
recognised quickly and rectified using the following table. If you
cannot find the error in the table or the problem cannot be
rectified using the method described, please contact your
specialist retail partner.
You will reach your authorised specialist retail partner via
contact form on the manufacturer’s website
www.geratherm-respiratory.com/login/
Danger to life.
Reason: Unauthorised work carried out by the user to
troubleshoot and rectify an error. Therefore:
•
Users may only carry out work which is
described and permitted by the manufacturer!
In case of not complying with this restriction, the
responsible operator alone is liable for any resulting
injuries to persons and / or damages to Spirostik!
Error
Rectification
Software error message:
„Flow out of range”
Perform zero adjustment.
Check the tube connection between handle and
Spirostik.
Disconnect Spirostik from PC and reconnect to
USB port.
Software error message:
“Internal communication
error”
Disconnect Spirostik from PC and reconnect to
USB port.
obvious measuring error Check last calibration date and recalibrate if
necessary (see chap. 7.5)