12.2
Installation and Operating Conditions
The following conditions supplement chap. 2.2 "Intended Use"
and must be observed to maintain the properties of the Spirostik
Complete guaranteed by the manufacturer and for the safety of
the patient and operator.
Storage / Transport:
min.
max.
Temperature:
-10
C
+50
C
Relative air humidity:
0
%
95
%
non-condensing
Operation:
min.
max.
Environmental temperature:
+10
C
+35
C
Avoid extreme
fluctuations in
temperature!
Relative air humidity:
20
%
95
%
non-condensing
(min. 30
% for synthetic flooring)
Atmospheric pressure:
800
hPa
1080
hPa
Setup:
mobile
Space requirement:
(L) 330
mm x (W) 355
mm x (H) 75
mm
Flooring:
stable; even; conductive
no synthetic material if possible
Environment / room:
Closed, clinical area
air-conditioned
not an explosive or flammable environment!