Instructions for Use
Bodystik
Serial no.: xx|8|301|yyy and 2301xxxxx
Version: 7 | Release date: 02. August 2021
Page 71
7.2
Switching Bodystik on
The Bodystik is designed for permanent operation and doesn't
need to be switched on and off daily.
But there are different ways to finish the work with the Bodystik.
Therefore the Bodystik is "switched on" again in different ways.
For descriptions of "Recommissioning after Servicing / Cleaning
Work" see:
•
•
Chap. 6.2.1 "Recommissioning after Termination by
Patient"
•
Chap. 6.2.2 "Recommissioning after Maintenance /
Cleaning Work by the Operator"
7.3
Calibrating Bodystik
There are various processes to be carried out for Bodystik in
order to calibrate or validate the volume measurement.
The required intervals can be found in the tables below.
Interval
Calibration
Once per day
Box calibration
Once per day, and after changing the Flow
Sensor
Volume calibration
Once per week
Flow linearity test
Once per year
Mouth pressure calibration
For details of individual calibrations, please read the separate
IFU:
•
Flow Sensor Calibration
•
Calibration Bodyplethysmograph