Instructions for Use
Spirostik Blue
Serial no.: 2111xxxxx
Version: 7 | Release date: 02. August 2021
Page 81
12.2
Installation and Operating Conditions
The following conditions supplement chap. 2.2 "Intended Use"
and must be observed to maintain the properties of the Spirostik
Blue guaranteed by the manufacturer and for the safety of the
patient and operator.
Storage / Transport:
min.
max.
Temperature:
-10 °C
+50 °C
Relative air humidity:
20 %
95 %
non-condensing
Atmospheric pressure:
700 hPa
1100 hPa
Operation:
min.
max.
Environmental temperature:
+15 °C
+35 °C
Avoid extreme
fluctuations in
temperature!
Relative air humidity:
20 %
95 %
non-condensing
(min. 30 % for synthetic flooring)
Atmospheric pressure:
700 hPa
1080 hPa
Setup:
mobile
Space requirement:
(L) 210 mm x (W) 130 mm x (H) 50 mm
Environment / room:
Closed, clinical area
air-conditioned
not an explosive or flammable environment!
not near an MRI!
well ventilated
Operating Mode:
Continuous operation