TEMPHION INSTRUCTIONS

Seametrics • 253.872.0284 

Page 23 seametrics.com

Battery Changing Tips

•  Never place a tool on the sensor body, it is very 

thin and will deform causing leaks at o-ring seals 

and potentially crushing the circuit board!

• 

Always twist the sensor body off the top cap 

assembly rather than twisting the top cap assembly 

off of the sensor body.

•  For cabled sensors, always clamp the sensor on the 

swaged area when applicable, the shoulder above 

it will allow you to press down without the worry of 

the sensor slipping out of the clamping device.

•  If the sensor body is slippery or you are unable to 

grip it hard enough to twist, try a piece of rubber 

cabinet liner for additional friction.

There is a black, compressible square ring near the top of 

the sensor. This ring acts as a spring to lock the cable in. 

This needs to be compressed in order to allow removal 

of the top cap. Once this ring is compressed, a gentle 

counterclockwise twist is all that is needed to remove the 

cable from the sensor body. Compressing the black square 

ring does take force, 

twisting does not.

Cabled Sensor

Cableless Sensor

Care must be taken to compress the black square ring 

before attempting to twist the housing. 

Forceful twisting 

of the housing can permanently damage the sensor

.

Securing the Sensor

In order to compress the black square ring, the sensor 

must be secured so that you can apply downward pressure 

to compress the ring. This can be done by holding in your 

hand, using a vise, or using pliers, as follows.

Swage Knurling Black square ring

Housing

Top cap

Knurling Black square ring

Housing

Top cap

Battery Type:

 Two 1.5V AA  batteries—Lithium or Alkaline 

(lithium recommended). The TempHion sensors are 

shipped from the factory with fresh batteries. Battery life 

will depend on battery brand, battery age, temperature of 

the environment, usage schedule, and other factors.

IMPORTANT!

Because changing the batteries involves 

opening the water-tight seal, this must be 

done in a clean, dry environment to avoid 

contamination or moisture damage to the 

circuitry.

IF USING ALKALINE BATTERIES 

—PREVENT BATTERY LEAKAGE!

TempHion sensors are typically shipped with 

lithium batteries. If, however, you are using 

alkaline batteries, be aware that under some 

circumstances alkaline batteries can leak, 

causing damage to the sensor. To prevent 

leakage, the following is recommended. 

(Does 

not apply to lithium batteries.)

•  Change the batteries at least every 18 

months.

•  If the sensor will not be deployed for 3 

months or more, remove the batteries.

Changing Batteries

MAINTENANCE and REFERENCE

Reconditioning 

Reference Electrode

• 

If stored dry or in KCl, refill per maintenance section 

earlier in this manual. Once refilled,  let sit for 30 

minutes before using.

•  If stored in reference solution, no reconditioning is 

needed. 

pH bulb / ORP

• 

If stored dry, soak in 1st buffer for 4 hours before 

calibrating. (Must be calibrated before use.)

•  If stored in KCl or pH 4, no reconditioning is 

needed. 

ISE

•  If stored dry, soak in 1st calibration solution for 30 

minutes before calibrating.  (Must be calibrated 

before use.)

Содержание TempHion ISE

Страница 1: ...SA C e rtified Compa n y ISO 9001 2008 TempHion pH ORP Redox ISE Smart Sensor and Data Logger Instructions TempHion pH ORP Redox TempHion ISE For Sales Service Contact 2650 E 40th Ave Denver CO 80205...

Страница 2: ...TEMPHION INSTRUCTIONS Seametrics 253 872 0284 Page 2 seametrics com...

Страница 3: ...ng Page 16 Retrieving Data Page 16 Viewing Data Page 16 Exporting Data Page 17 A Word about Units Page 17 Direct Read Modbus SDI 12 Setting Units for Direct Read Page 17 Power Consideration Page 18 Re...

Страница 4: ...Alternately it can be powered with an external auxiliary power supply for data intensive applications The unit is programmed using Seametrics easy to use Aqua4Plus or Aqua4Plus Lite control software...

Страница 5: ...s a4d and csv also xls in Windows 8 and earlier Output Channels Temperature pH ORP Bromide Element 30K ohm thermistor Epoxy bead external housing Glass combination electrode Platinum ring Ag AgCl soli...

Страница 6: ...ge is found immediately file a claim with the carrier Check the label attached to the cable at the connector end for the proper cable length Do s and Don ts Don t drop into well Lower gently to preven...

Страница 7: ...ed up for cleaning refilling or other maintenance Reference Electrode Assembly The modified thread provides a continuous liquid path between the solution in which the instrument is immersed and the re...

Страница 8: ...nal batteries If auxiliary power is desired you can use a 6 16 VDC supply that can provide 15 mA Connect to Vaux pin 1 white and Ground pin 5 blue or contact Seametrics for auxiliary power supplies Co...

Страница 9: ...rics Aqua4Plus or Aqua4Plus Lite software However the TempHion is quite versatile communicating via either Modbus or SDI 12 interfaces allowing you to do the following Read a TempHion via Modbus using...

Страница 10: ...t at all times reach high enough to cover the sensing element cutout Be sure the supplied cap is securely placed on the weather resistant connector at the top of the cable Do not install such that the...

Страница 11: ...he window Notethatthetemperaturechannelrarelyneedsadjustment If you think your temperature channel needs calibrating contact your service representative Use the calibration calculator to perform one a...

Страница 12: ...ance and Reference section From this chart determine the pH value at your current buffer temperature and use this value as the Ref pH value in the calibration instructions below For example if you are...

Страница 13: ...roughout the calibration Temperature of the sensor must also be this temperature prior to calibration One Point Calibration Computing Calibration Value Rinse sensor with distilled water and pat dry Pl...

Страница 14: ...Point Add 10 cc of the same standard to the water Depending on which solution you are using this will result in a concentration as shown below 0 1 Molar NaBr Bromide 791 8 ppm Stir to distribute stan...

Страница 15: ...nel do not use the built in calculator in the Settings and Calibration Window Instead follow the instructions below Rinse sensor with distilled or deionized water and then pat dry with clean paper tow...

Страница 16: ...of the current readings on the sensor Setting up Data Recording Click the tool button A Logging Schedule Window will open Refer to the software instruction manuals for details in describing your sched...

Страница 17: ...quipment or networks You may need to use Aqua4Plus Aqua4Plus Lite to make a few settings prior to directly reading the TempHion with your equipment These might include the units for the returned value...

Страница 18: ...Holding Registers Readings are located in two registers each starting at address 62592 TempHion register addressing is zero based i e starts at zero If your equipment uses one based addressing you wi...

Страница 19: ...for temperature is Celsius To change this use the Direct Read Units option in the Aqua4Plus Aqua4Plus Lite Control Software When using the M MC C or CC command all mV channels will report in either p...

Страница 20: ...n mV Concurrent Measurement a C a 0024 cr lf Request all measurements a D0 a 21 345 7 181 053 459 cr lf Read temperature pH in pH units ISE in ppm ORP in Eh a C1 a 0024 cr lf Request all measurements...

Страница 21: ...Seametrics 0 54 0 Chloride ION Plus B 3 572 0 49 Bromide Seametrics 3 878 0 89 CAUTION Filling solutions are not considered hazardous but they can be irritating to the skin Protective gloves are advi...

Страница 22: ...below when you calibrate the sensor To determine which configuration you have you will need to create a short data file then retrieve and export the data it doesn t matter whether the sensor is deploy...

Страница 23: ...an be done by holding in your hand using a vise or using pliers as follows Swage Knurling Black square ring Housing Top cap Knurling Black square ring Housing Top cap Battery Type Two 1 5V AA batterie...

Страница 24: ...in next page Cableless Sensor By Hand cabled version only 1 Tightly grasp the top cap in one hand 2 Brace your hand against something such as a table or the ground Do not allow the cable to be pinched...

Страница 25: ...onnector it only goes in one way so make sure not to force it Pull wires gently to the side to allow battery removal Connector connected properly 4 Hold the top cap assembly at 90 to the housing openi...

Страница 26: ...S BELOW Incorrect readings O ring has not been moved to expose the reference reservoir opening Move the O ring from the upper groove to the lower groove See the Installation section Element or referen...

Страница 27: ...ltered or repaired by the Buyer without prior written approval by the Seller all warranties are void Equipment and accessories not manufactured by the Seller are warranted only to the extent of and by...

Страница 28: ...Seametrics 19026 72nd Avenue South Kent Washington 98032 USA P 253 872 0284 F 253 872 0285 1 800 975 8153 seametrics com LT 14315r14 20180306 3 6 2018...

Отзывы: