TempHion

 

pH/ORP (Redox) / ISE Smart Sensor  

and Data Logger Instructions

TempHion pH/ 

ORP (Redox)

TempHion ISE

For Sales & Service Contact

2650 E. 40th Ave. • Denver, CO 80205

Phone 303-320-4764 • Fax 303-322-7242

1-800-833-7958

www.geotechenv.com

Содержание TempHion ISE

Страница 1: ...SA C e rtified Compa n y ISO 9001 2008 TempHion pH ORP Redox ISE Smart Sensor and Data Logger Instructions TempHion pH ORP Redox TempHion ISE For Sales Service Contact 2650 E 40th Ave Denver CO 80205...

Страница 2: ...TEMPHION INSTRUCTIONS Seametrics 253 872 0284 Page 2 seametrics com...

Страница 3: ...ng Page 16 Retrieving Data Page 16 Viewing Data Page 16 Exporting Data Page 17 A Word about Units Page 17 Direct Read Modbus SDI 12 Setting Units for Direct Read Page 17 Power Consideration Page 18 Re...

Страница 4: ...Alternately it can be powered with an external auxiliary power supply for data intensive applications The unit is programmed using Seametrics easy to use Aqua4Plus or Aqua4Plus Lite control software...

Страница 5: ...s a4d and csv also xls in Windows 8 and earlier Output Channels Temperature pH ORP Bromide Element 30K ohm thermistor Epoxy bead external housing Glass combination electrode Platinum ring Ag AgCl soli...

Страница 6: ...ge is found immediately file a claim with the carrier Check the label attached to the cable at the connector end for the proper cable length Do s and Don ts Don t drop into well Lower gently to preven...

Страница 7: ...ed up for cleaning refilling or other maintenance Reference Electrode Assembly The modified thread provides a continuous liquid path between the solution in which the instrument is immersed and the re...

Страница 8: ...nal batteries If auxiliary power is desired you can use a 6 16 VDC supply that can provide 15 mA Connect to Vaux pin 1 white and Ground pin 5 blue or contact Seametrics for auxiliary power supplies Co...

Страница 9: ...rics Aqua4Plus or Aqua4Plus Lite software However the TempHion is quite versatile communicating via either Modbus or SDI 12 interfaces allowing you to do the following Read a TempHion via Modbus using...

Страница 10: ...t at all times reach high enough to cover the sensing element cutout Be sure the supplied cap is securely placed on the weather resistant connector at the top of the cable Do not install such that the...

Страница 11: ...he window Notethatthetemperaturechannelrarelyneedsadjustment If you think your temperature channel needs calibrating contact your service representative Use the calibration calculator to perform one a...

Страница 12: ...ance and Reference section From this chart determine the pH value at your current buffer temperature and use this value as the Ref pH value in the calibration instructions below For example if you are...

Страница 13: ...roughout the calibration Temperature of the sensor must also be this temperature prior to calibration One Point Calibration Computing Calibration Value Rinse sensor with distilled water and pat dry Pl...

Страница 14: ...Point Add 10 cc of the same standard to the water Depending on which solution you are using this will result in a concentration as shown below 0 1 Molar NaBr Bromide 791 8 ppm Stir to distribute stan...

Страница 15: ...nel do not use the built in calculator in the Settings and Calibration Window Instead follow the instructions below Rinse sensor with distilled or deionized water and then pat dry with clean paper tow...

Страница 16: ...of the current readings on the sensor Setting up Data Recording Click the tool button A Logging Schedule Window will open Refer to the software instruction manuals for details in describing your sched...

Страница 17: ...quipment or networks You may need to use Aqua4Plus Aqua4Plus Lite to make a few settings prior to directly reading the TempHion with your equipment These might include the units for the returned value...

Страница 18: ...Holding Registers Readings are located in two registers each starting at address 62592 TempHion register addressing is zero based i e starts at zero If your equipment uses one based addressing you wi...

Страница 19: ...for temperature is Celsius To change this use the Direct Read Units option in the Aqua4Plus Aqua4Plus Lite Control Software When using the M MC C or CC command all mV channels will report in either p...

Страница 20: ...n mV Concurrent Measurement a C a 0024 cr lf Request all measurements a D0 a 21 345 7 181 053 459 cr lf Read temperature pH in pH units ISE in ppm ORP in Eh a C1 a 0024 cr lf Request all measurements...

Страница 21: ...Seametrics 0 54 0 Chloride ION Plus B 3 572 0 49 Bromide Seametrics 3 878 0 89 CAUTION Filling solutions are not considered hazardous but they can be irritating to the skin Protective gloves are advi...

Страница 22: ...below when you calibrate the sensor To determine which configuration you have you will need to create a short data file then retrieve and export the data it doesn t matter whether the sensor is deploy...

Страница 23: ...an be done by holding in your hand using a vise or using pliers as follows Swage Knurling Black square ring Housing Top cap Knurling Black square ring Housing Top cap Battery Type Two 1 5V AA batterie...

Страница 24: ...in next page Cableless Sensor By Hand cabled version only 1 Tightly grasp the top cap in one hand 2 Brace your hand against something such as a table or the ground Do not allow the cable to be pinched...

Страница 25: ...onnector it only goes in one way so make sure not to force it Pull wires gently to the side to allow battery removal Connector connected properly 4 Hold the top cap assembly at 90 to the housing openi...

Страница 26: ...S BELOW Incorrect readings O ring has not been moved to expose the reference reservoir opening Move the O ring from the upper groove to the lower groove See the Installation section Element or referen...

Страница 27: ...ltered or repaired by the Buyer without prior written approval by the Seller all warranties are void Equipment and accessories not manufactured by the Seller are warranted only to the extent of and by...

Страница 28: ...Seametrics 19026 72nd Avenue South Kent Washington 98032 USA P 253 872 0284 F 253 872 0285 1 800 975 8153 seametrics com LT 14315r14 20180306 3 6 2018...