Installation

Fitting Installation

•  Cut hole in pipe at desired location—recommended 

size is 1 3/4”. 

• 

Measure and write down the wall thickness.

•  Strap saddle on pipe or weld the weldolet to the 

pipe, aligning over hole.

Determine Proper Depth and K-Factor

•  Run the 

K-factor Calculator

, located at the bottom of 

the seametrics.com web site to obtain Depth Setting 

“D” and the K-factor.

Meter Installation

•  Loosen compression nut so adapter shaft slides 

freely. Pull meter fully upward and finger tighten 

compression nut.

• 

Install the adapter in the pipe fitting, using a thread 

sealant.

Positioning & Grounding

Choose a position that will ensure a full pipe. 

Choose a position that will minimize flow distortion. 

10X

5X

10X

5X

20X

Fair — Unacceptable if pipe  

 

contains air

Best Position

Fair — Unacceptable if pipe  

 

contains sediment

Connect grounding wire from 

one of the housing screws  or the 

shaft of meter to a good earth 

ground. If the pipe is metal, also 

connect the grounding wire to a 

grounding clamp on the pipe.

Metallic Pipe

Earth Ground

Grounding  

Clamp

Housing Screw

• 

Reinstall the sensor unit, being sure flow is in the 

direction of the arrow on the housing.

•  Loosen compression nut and lower meter to 

appropriate depth “D”, which was obtained using 

the K-factor calculator. 

•  Fully tighten compression nut.

“D”

Compression nut

Adapter mates with 

1.5” FNPT threads