Quick Start Guide
Seametrics • 19026 72nd Avenue South • Kent, Washington 98032 • USA
(P) 253.872.0284 • (F) 253.872.0285 • 1.800.975.8153 • seametrics.com
Installation
Fitting Installation
• Cut hole in pipe at desired location—recommended
size is 1 3/4”.
•
Measure and write down the wall thickness.
• Strap saddle on pipe or weld the weldolet to the
pipe, aligning over hole.
Determine Proper Depth and K-Factor
• Run the
K-factor Calculator
, located at the bottom of
the seametrics.com web site to obtain Depth Setting
“D” and the K-factor.
Meter Installation
• Loosen compression nut so adapter shaft slides
freely. Pull meter fully upward and finger tighten
compression nut.
•
Install the adapter in the pipe fitting, using a thread
sealant.
Positioning & Grounding
Choose a position that will ensure a full pipe.
Choose a position that will minimize flow distortion.
10X
5X
10X
5X
20X
Fair — Unacceptable if pipe
contains air
Best Position
Fair — Unacceptable if pipe
contains sediment
Connect grounding wire from
one of the housing screws or the
shaft of meter to a good earth
ground. If the pipe is metal, also
connect the grounding wire to a
grounding clamp on the pipe.
Metallic Pipe
Earth Ground
Grounding
Clamp
Housing Screw
•
Reinstall the sensor unit, being sure flow is in the
direction of the arrow on the housing.
• Loosen compression nut and lower meter to
appropriate depth “D”, which was obtained using
the K-factor calculator.
• Fully tighten compression nut.
“D”
Compression nut
Adapter mates with
1.5” FNPT threads