Keck 1500’ – 2000’ ET/Tuff Tape
Water Level Meters
with optional electric rewind
Installation and operation Manual
Страница 1: ...Rev 2 09 09 09 Part 12050339 Keck 1500 2000 ET Tuff Tape Water Level Meters with optional electric rewind Installation and operation Manual...
Страница 2: ......
Страница 3: ...NSTALLATION 6 MANUAL REEL 6 ELECTRIC REEL 6 CHAPTER 3 SYSTEM OPERATION 7 CHAPTER 4 SYSTEM MAINTENANCE 8 CHAPTER 5 SYSTEM TROUBLESHOOTING 9 CHAPTER 6 SYSTEM SPECIFICATIONS 11 CHAPTER 7 SYSTEM SCHEMATIC...
Страница 4: ...ceed until you read and thoroughly understand the WARNING message WARNING CAUTION A raised hand icon indicates CAUTION information that relates to a situation or condition that could lead to equipment...
Страница 5: ...O D stainless steel weighted probe attached to a polyethylene tape marked to 1 100th of a foot The meter relies on fluid conductivity to determine the presence of water An audible signal and visible L...
Страница 6: ...4 System Components Figure 1 Manual Reel Front View Figure 2 Manual Reel Side View Visible Signal Power On Sensitivity switch Reel Handle Engineers tape Frame Pin Lock Brake 5 8 Probe...
Страница 7: ...5 Figure 3 Electric Reel Front View Figure 4 Electric Reel Side View Visible Signal Power On Sensitivity Switch Probe Battery Clips Speed Control Up Down Push Buttons Frame Probe holder...
Страница 8: ...ositive battery terminal and black clip to the negative battery terminal 3 Turn the speed control dial fully counter clockwise Press desired button up down Slowly rotate the speed control until reel i...
Страница 9: ...still in contact with the water adjust the probe sensitivity dial clockwise until the light and buzzer barely activate In this setting the probe will detect water levels but will not be affected by fa...
Страница 10: ...an be cleaned with any detergent such as trisodium phosphate TSP Alconox or Liquinox If other detergents are used take care to select detergents that are compatible with Teflon polyethylene and stainl...
Страница 11: ...NO INDICATION OF WATER Solutions 1 The conductive contact is dirty Clean the contact see page 8 2 There is an open connection in the tape Replace tape and or probe 3 The circuit is malfunctioning Cons...
Страница 12: ...ctive contact is dirty causing bridging Clean the contact see page 8 2 There is a short in the tape and or probe Replace tape and or probe 3 The circuit is malfunctioning Consult the manufacturer For...
Страница 13: ...1500 450 m 50 4 lbs man 1500 450 m 60 5 lbs elec 2000 600 m 70 4 lbs man 2000 600m 80 5 lbs elec Tape accuracy 100th of a foot 100 Reel Frame Material Steel and Aluminum Size 18 25 H x 14 W x 15 L man...
Страница 14: ...12 Chapter 7 System Schematic Manual Reel Front and Side view Electric Reel Front and side view 14 10 16 5 15 20 10 14 16 5...
Страница 15: ...cription 52050144 Reel assy 1500 2000 manual 52050145 Reel assy 1500 2000 electric 52050124 Tape assy poly 1500 52050125 Tape assy poly 2000 52050248 Tape assy poly 450 m 52050249 Tape assy poly 600 m...
Страница 16: ...14 Notes...
Страница 17: ...15 Notes...
Страница 18: ...ial Authorization number RMA is required prior to return of any equipment to our facilities please call 800 number for appropriate location An RMA will be issued upon receipt of your request to return...
Страница 19: ......
Страница 20: ...ver Colorado 80205 303 320 4764 800 833 7958 FAX 303 322 7242 email sales geotechenv com website www geotechenv com In Michigan Call Geotech Environmental Equipment Inc 1099 West River Avenue Williams...