Section 2: System Installation 

Installing the High Vacuum Sampler 

1.

Verify that t

he power switch is in the “OFF” position.

2.

Plug the DC power input cord into a fuse protected 12V lighter receptacle on a
negatively grounded system or plug the optional AC power cord with 12VDC
adapter into a 110VAC power source. If the unit is wired directly to a 12V battery,
make sure to connect the positive (red) alligator clip to the positive (red) battery
terminal and connect the negative (black) alligator clip to the negative (black)
battery terminal.

3.

Securely plug the circular plug connector on the opposite end of the cable to the
corresponding receptacle on the face of the controller.

4.

Turn on the High Vacuum Sampler briefly and verify its operation. Place your
thumb over both ports: Air should be blowing out the AIR OUT port and vacuum
should be present at the VACUUM IN port. Use the System Troubleshooting
procedures in Section 5 when there is no air flow or vacuum response on the
gauge.

5.

Attach an air hose to the sample source and the other end to the VACUUM IN port
on the sampler. Next, connect an air hose to the AIR OUT port and the other end to
the sample bag.

The High Vacuum Sampler is ready to collect the sample. 

Verify the 12VDC system to be used is rated and fuse protected for 
operation at 15 amps continuous operation. Failure to verify system 
rating could result in damage to equipment. Over-heated wiring 
and other materials or components in contact or near by the power 
delivery system could pose a potential fire or burn hazard. 

Envirotecnics 

Tel: +34 872 080 542 

 

envirotecnics@envirote

cnics.com

www.envirotecnics.com

Содержание High Vacuum Sampler

Страница 1: ...Rev 05 06 13 Part 12151001 Geotech High Vacuum Sampler Installation and Operation Manual...

Страница 2: ...TEM OPERATION 7 SECTION 4 SYSTEM MAINTENANCE 8 SECTION 5 SYSTEM TROUBLESHOOTING 9 SECTION 6 SYSTEM SPECIFICATIONS 11 SECTION 7 SYSTEM SCHEMATIC 12 SECTION 8 REPLACEMENT PARTS AND ACCESSORIES 13 THE WA...

Страница 3: ...d the WARNING message WARNING CAUTION A raised hand icon indicates CAUTION information that relates to a situation or condition that could lead to equipment malfunction or damage You should not procee...

Страница 4: ...r Soil Vapor Extraction SVE systems Figure 1 1 is an example of the High Vacuum Sampler Each High Vacuum Sampler comes with a standard DC power cord Figure 1 2 to allow the unit to operate from a 12VD...

Страница 5: ...4 Figure 1 2 Figure 1 3 Example of ALTEF 1 Liter Bag Envirotecnics Tel 34 872 080 542 envirotecnics envirotecnics com www envirotecnics com...

Страница 6: ...5 Figure 1 4 Example of Tedlar 1 and 3 Liter Bags Envirotecnics Tel 34 872 080 542 envirotecnics envirotecnics com www envirotecnics com...

Страница 7: ...fly and verify its operation Place your thumb over both ports Air should be blowing out the AIR OUT port and vacuum should be present at the VACUUM IN port Use the System Troubleshooting procedures in...

Страница 8: ...the High Vacuum Sampler 2 Turn on the unit and allow the air flow to run prior to attaching your sample bag so that a good sample can be acquired 3 Push the valve stem in on the bag to open the valve...

Страница 9: ...ap and water solution can be applied to a soft cloth and used to clean the exterior surfaces of the High Vacuum Sampler Do not immerse soak or directly spray liquids on the High Vacuum Sampler Equipme...

Страница 10: ...ve The High Vacuum Sampler is protected from damage due to reverse polarity connection If ambient temperatures are in excess of 104 F 40 C then disconnect from power and let sit in a cool location Do...

Страница 11: ...be sent back to a factory authorized repair location Problem Compressor turns on and there is vacuum along with air flowing out but there is no vacuum indicated on the gauge Solutions The pressure gau...

Страница 12: ...W 17 02cm H x 12 92cm D x 28 96cm W Weight 11 4lbs 5 17kg Case optional 21 5 L x 15 H x 7 5 W 54 6 cm L x 38 1cm H x 19 05cm W Case Weight 7 9 lbs 3 6 kg total Free Air CFM 10 Vac in Hg 0 38 Free Air...

Страница 13: ...12 Section 7 System Schematic Figure 7 1 High Vacuum Sampler Wiring Diagram Envirotecnics Tel 34 872 080 542 envirotecnics envirotecnics com www envirotecnics com...

Страница 14: ...NCLUDES 10FT TUBING 57500008 ASSY POWER CORD DC W AMP 17500035 ADAPTER CIGARETTE TO CLIPS 77250000 BATTERY CHARGER MODULAR 12VDC 8Ah MALE CIG PLUG 51150127 CASE GEOCONTROL PRO CE 12150020 TEDLAR BAG 3...

Страница 15: ...F DESCRIPTION REV DATE Previous Release 12 17 12 1570 Edited description for spare part 87050509 Added Revision History Table SP 05 06 13 Envirotecnics Tel 34 872 080 542 envirotecnics envirotecnics c...

Страница 16: ...or to return of any equipment to our facilities please call our 800 number for appropriate location An RMA will be issued upon receipt of your request to return equipment which should include reasons...

Страница 17: ...7958 FAX 303 322 7242 Envirotecnics Global Service SL Oficinas y Fabricaci n Calle dels Pirineus s n esquina Espinau 17460 Celr Girona Tel 872 080 542 Fax 872 080 543 Delegaci n Madrid Avda de Castill...

Отзывы: