User Manual - P series
23
www.georgin.com
»
Year 12 :
After 12 years in standard operation, Georgin recommends sending the instrument to our After Sale
Service department for diagnosis. This may lead to some components being replaced or the instrument
being scrapped.
After 12 years of operation, an annual inspection must be performed.
Note: Beyond 20 years of operation, no further repairs will be carried out. The instrument will always be
scrapped in the event of failure.
*In view of the many and varied possible uses on industrial sites, the maintenance frequency must be
increased in case of heavy-duty operation or harsh conditions.
For example:
-
High rate > 4 cycles/minute
-
Corrosive or crystalline fluid or environments
!
Instruments in safety roles must be inspected annually.
2. Precautions to take during maintenance
The instrument must be disassembled WHEN DISCONNECTED FROM POWER, using an appropriate spanner for the
fitting.
GEORGIN guarantees the qualification of the equipment that leaves the factory. Any operation performed on
the equipment other than adjusting the instrument releases GEORGIN from all liability in the event of failure.
In the event you suspect a failure or permanent fault, return the equipment to our offices or our
representatives, who are the only maintenance providers certified to perform expert assessments or repairs.
Any equipment returned MUSTbe accompanied by an After Sale Service return form. This can be obtained
from www.georgin.com (look under download/After Sale Service) and must be duly completed and signed.