Geokon 3800 Скачать руководство пользователя страница 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No part of this instruction manual may be reproduced, by any means, without the written consent of Geokon, Inc. 

The information contained herein is believed to be accurate and reliable.  However, Geokon, Inc. assumes no responsibility for 

errors, omissions or misinterpretation.  The information herein is subject to change without notification. 

Copyright ©, 2009-2016 by Geokon, Inc. 

(Doc Rev E, 04/07/17) 

Installation Manual 

Model 3800/3810 

Thermistors & Thermistor Strings 

 

Содержание 3800

Страница 1: ...ontained herein is believed to be accurate and reliable However Geokon Inc assumes no responsibility for errors omissions or misinterpretation The information herein is subject to change without notification Copyright 2009 2016 by Geokon Inc Doc Rev E 04 07 17 Installation Manual Model 3800 3810 Thermistors Thermistor Strings ...

Страница 2: ......

Страница 3: ...o warranties except as stated herein There are no other warranties expressed or implied including but not limited to the implied warranties of merchantability and of fitness for a particular purpose Geokon Inc is not responsible for any damages or losses caused to other equipment whether direct indirect incidental special or consequential which the purchaser may experience as a result of the insta...

Страница 4: ......

Страница 5: ...TION 1 2 INSTALLATION 1 3 READOUT 1 4 DATALOGGER CONNECTION 1 FIGURE A 1 THERMISTOR BRIDGE WIRING 2 TABLE B 1 STANDARD THERMISTOR RESISTANCE VERSUS TEMPERATURE 4 TABLE B 2 HIGH TEMPERATURE THERMISTOR RESISTANCE VERSUS TEMPERATURE 6 ...

Страница 6: ......

Страница 7: ...Model 3800 1 1 1 into one of the pairs of conductors then waterproofing the splice Before Installation read and record all the thermistors to check that they are functional and to establish ambient temperature readings Read and record once again as soon as the thermistor string has been deployed and again after a certain length of time has elapsed to allow the temperatures stabilize so as to estab...

Страница 8: ...put G G Power Ground C RES Control Port Reset C CLK Digital I O Port Clock AG AG Analog Ground SE 1H Single Ended CH input for Thermistor VX 1L Excitation VX NOTE The Thermistors are read using a Single Ended Channel and Excited with one of the dataloggers excitation channels through a bridge completion circuit Figure A 1 Thermistor Bridge Wiring ...

Страница 9: ...e Each individual Thermistor is connected with common jumper as shown below Multiplexer Wiring 8032 Multiplexer Model 3800 3810 IH Thermistor Gage 1 1L Thermistor Common 2H Thermistor Gage 2 2L Thermistor Common S Shield 3H Thermistor Gage 3 3L Thermistor Common NOTE Jumper required ...

Страница 10: ...8417 3 1418 43 342 2 83 107 9 123 77 66K 36 8006 4 1363 44 331 5 84 105 2 124 72 81K 35 7618 5 1310 45 321 2 85 102 5 125 68 30K 34 7252 6 1260 46 311 3 86 99 9 126 64 09K 33 6905 7 1212 47 301 7 87 97 3 127 60 17K 32 6576 8 1167 48 292 4 88 94 9 128 56 51K 31 6265 9 1123 49 283 5 89 92 5 129 53 10K 30 5971 10 1081 50 274 9 90 90 2 130 49 91K 29 5692 11 1040 51 266 6 91 87 9 131 46 94K 28 5427 12 ...

Страница 11: ...8 2 6 808 34 2 011 66 719 6 98 301 0 130 142 0 162 73 6 194 41 4 226 28 052 3 6 531 35 1 942 67 698 7 99 293 5 131 138 9 163 72 2 195 40 7 227 26 685 4 6 267 36 1 876 68 678 6 100 286 3 132 135 9 164 70 8 196 40 0 228 25 392 5 6 015 37 1 813 69 659 1 101 279 2 133 133 0 165 69 5 197 39 3 229 24 170 6 5 775 38 1 752 70 640 3 102 272 4 134 130 1 166 68 2 198 38 7 230 23 013 7 5 545 39 1 693 71 622 2...

Страница 12: ...81 51 8 213 30 4 245 11 419 22 3 099 54 1 037 86 410 8 118 186 1 150 93 2 182 50 9 214 29 9 246 10 922 23 2 986 55 1 005 87 400 0 119 181 9 151 91 3 183 50 0 215 29 4 247 10 450 24 2 878 56 973 8 88 389 6 120 177 7 152 89 5 184 49 1 216 29 0 248 10 000 25 2 774 57 944 1 89 379 4 121 173 7 153 87 7 185 48 3 217 28 5 249 9 572 26 2 675 58 915 5 90 369 6 122 169 8 154 86 0 186 47 4 218 28 1 250 9 165...