Counter and Timer Control Feature Description
, describes these parameters along with their view attribute
and minimum camera firmware version required. Additionally the Device Version column will
indicate which feature is a member of the DALSA Features Naming Convention (indicated by
DFNC), versus the GenICam Standard Features Naming Convention (SFNC tag is not shown).
The Device Version number represents the camera software functional group, not a firmware
revision number. As Nano-CL capabilities evolve the device version tag will increase, therefore
identifying the supported function package. New features for a major device version release will be
indicated by
green text
for easy identification.
Display Name
Feature & Values
Description
Device
Version
& View
Counter Selector
counterSelector
Selects the counter to configure.
1.00
Expert
DFNC
Counter 1
Counter1 Select counter 1
Counter mode
counterMode
Selects the counter mode. The selected Counter is
either Active or Disabled. When Disabled, the Counter
can be configured.
1.00
Expert
DFNC
Off
Off The selected Counter is Disabled
Active
Active The selected Counter is Enabled
Counter Status
counterStatus
Returns the current state of the counter.
1.00
Expert
DFNC
Counter Idle
CounterIdle The counter is idle.
The counterStartSource feature is set to off.
Counter Trigger Wait
CounterTriggerWait The counter is waiting for a start trigger.
Counter Active
CounterActive The counter is counting for the specified duration.
Counter Completed
CounterCompleted The counter reached the CounterDuration count.
Counter Overflow
CounterOverflow The counter reached its maximum possible count.
counterStartSource
Select the counter start source. Counter increments
from 0 to the value of the counterDuration feature.
1.00
Expert
DFNC
Off
Off Counter is stopped.
Exposure Start
ExposureStart Counter starts on the reception of the Exposure Start
event
Exposure End
ExposureEnd Counter starts on the reception of the Exposure End
event.
Readout Start
ReadoutStart Counter starts on the reception of the Readout Start
event.
Readout End
ReadoutEnd Counter starts on the reception of the Readout End
event.
Frame Start
FrameStart Counter starts on the reception of the Frame Start
event.
Valid Frame Trigger
ValidFrameTrigger Counter starts on the reception of the Valid Frame
Trigger.
Rejected Frame Trigger
InvalidFrameTrigger Counter starts on the reception of the Invalid Frame
Trigger.
CC1
CC1 Counter starts on the specified transitions on CC1.
Timer 1 End
Timer1End Counter starts on the reception of the Timer 1 End
event.
Counter 1 End
Counter1End Counter starts on the reception of the Counter 1 End
event.
Counter Start Line
Activation
counterStartLineActivation
Selects the activation mode of the input line trigger
which starts the counter. This is only applicable when
the counterStartSource feature selects a physical
Line.
1.00
Expert
DFNC
Содержание Nano-CL C2450
Страница 94: ...92 Technical Specifications Genie Nano CL Series EC FCC Declarations of Conformity...
Страница 104: ...102 Additional Reference Information Genie Nano CL Series Cable Assembly G3 AIOC BLUNT1M...
Страница 105: ...Genie Nano CL Series Additional Reference Information 103 Cable Assembly G3 AIOC BLUNT2M...
Страница 106: ...104 Additional Reference Information Genie Nano CL Series...
Страница 107: ...Genie Nano CL Series Additional Reference Information 105 Cable Assembly G3 AIOC BRKOUT2M...
Страница 108: ...106 Additional Reference Information Genie Nano CL Series...
Страница 109: ...Genie Nano CL Series Additional Reference Information 107 Nano Generic Power Supply with no I O...
Страница 111: ...Genie Nano CL Series Additional Reference Information 109 Cable Assembly Right Angle I O to Euro Block...