Genevac HT-24
User Manual
30
10-1562 Issue 3-7 March
2014
Data transfer
Data can be transferred to or from the evaporator by inserting a USB data key (not supplied)
into the USB A port, to the left of the control panel.
Insert the data key before switching the evaporator on. If necessary, switch the evaporator
off, wait ten seconds, insert the data key and switch the evaporator on again.
To transfer data, go to the
OPTIONS
menu and select
FILE TRANSFER
, then select one of
the following options:
COPY ALL METHODS FROM DATAKEY:
All methods on the data key will be
transferred to the evaporator (run slot 1 to 100). Where blank method slots exist on the
data key, evaporator methods will remain unchanged, all other methods will be
overwritten.
COPY ALL METHODS TO DATAKEY:
Methods can be copied for safekeeping or for
cloning another system
COPY LOG FILES TO DATAKEY:
Log files for the last four evaporation runs can be
downloaded for method validation, troubleshooting or method development purposes
SAVE SYSTEM INFO TO DATAKEY:
For troubleshooting use by Genevac Service
department
COPY SETUP.INI FROM DATAKEY:
For use by Genevac Service Engineers.
Confirmation requests or advice messages may appear on the LCD screen and file transfer
progress is indicated. When files are transferred to a data key for the first time, a sub-
directory, entitled
Series2
is created on the data key.
Note:
For reliable operation, use a small USB data key (4 GB or less) that does not have
protection, encryption or drive partition.