10.3.6
Trigger qualifier
The trigger detectors of a channel can also be used as qualifiers. A trigger
qualifier is a situation that enables (arms) the recorder trigger features. The
recorder trigger features are a combination of various channel, external,
between
-
recorders and other trigger options.
There are two qualifier modes:
l
Basic single
-
level qualifier. The level detector operates identical to "Basic
l
Dual
-
level qualifier. The level detector operates identical to "Dual trigger
When in qualifier mode, the output of the trigger detector is sent to a qualifier
line of the recorder trigger logic. For more information on the recorder trigger
features, please refer to "Recorder and system trigger" on page 152.