M02 Digital Indicator
13
3 Calibration
3.1 Instruction
(
1
)
Calibration procedure must be executed when a M02 indicator is put in use at the
first time, the preset parameters may no longer meet the user’s needs, and any part of
the weighing system was changed. Position of decimal point, minimum division,
maximum capacity, zero, and gain can be set and confirmed through calibration.
(
2
)
If you want to set only one parameter, please press
to save parameter’s
value and then press
to exit.
(
3
)
Please see section
3.7
for parameters’ instruction.
(
4
)
Please record each value in the blank table in section 3.4 during calibration for the
emergency use in future.
(
5
)
See chapter
9
for error alarm message that may be displayed during calibration.