GM7702 User Manual
- 8 -
1
. As load cell output signal is sensitive to electronic noise, the wiring should be shield cable
and separate with other cables, especially AC power.
2
.
4 wires connection is suitable for short distance and stable temperature or low precision
field, otherwise use 6 wires connection.
3
.
For more load cells parallel connection, their sensitivity (mV/V) should be the same.
2.3 Connection of I/O Terminal
GM7702
weighing indicator has equipped with I/O module of 2 IN and 3 OUT. IN/OUT
function user can self-define. To facilitate users wiring and some special applications, the
default definition is as follows
Out p ut
In pu t
OU T 1
s ta b le
IN 1
z ero
OU T 2
o ve rl i m it
IN 2
tare
OU T 3
set point 1