Mechanical Separation /
GEA Westfalia Separator
4.10 Alarm #12 - Vibration stage 2 (option)
VIBRATION STAGE 2:
Separator “Off”
Alarm#12 ENTER
Text display
Possible causes
Action
Limit value 2 set too low.
•
Check setting.
Uneven solid desposits in the bowl.
•
Check solids space and polyamide
gasket of the bowl top.
Vibro sensor has tripped.
•
Check sensor fastening.
Bowl not correctly mounted, damaged
or wrong parts from other bowl.
•
Open and clean the bowl and mount
in accordance with the manual.
CAUTION!
•
Use only parts with identical
serial number.
External vibrations
•
Check foundation fastening ar-
rangement.
Help text – information on the control system
E N T E R
Step 1
Acknowledge fault
H E L P
Step 2
Press key >F1<
Help on alarm
Key >HOME< back
Page 1
VIBRATION STAGE 2:
Please check:
Limit value 2
Solids space
->
Page 2
Polyamide gasket
Sensor fastening
Correct bowl
parts
->
Page 3
Foundation fastening
Back with ENTER
F 1