3.
Disconnect the power supply.
®
Disassembly has been prepared.
10.6
Disassembling and Assembling the Valve
10.6.1 Disassembling the valve
Tools required:
•
Hex socket key
Caution!
Danger of injury due to spring force being released
You can sustain injuries to your fingers when you put your hand into a valve
if the valve has not been moved to the open position beforehand.
► Before starting any work, bring the valve to the "open" position.
► Wear protective gloves for all work.
► Always exercise caution and prudence.
Caution!
Danger of injury due to media escaping after removing the clamp
You can sustain injuries to your entire body if you open a valve that is still
under pressure from the medium.
► Ensure that the valve is no longer under medium pressure before
removing the clamp.
► Wear suitable protective clothing for all work.
► Always exercise caution and prudence.
Carry out the following steps:
1.
Bring the valve to the "open" position.
Fig.31:
Apply valve NC
2.
Release the clamp with a suitable
hex socket key
, but do not unhook the
screw yet.
Maintenance
Disassembling and Assembling the Valve
430BAL011547EN_2
38
08.07.2019