Section 7-6 - Common Service Desktop (CSD)
7-6-5-1
InSite Configuration
1.) Enter the Common Service Desktop (CSD).
2.) Select the Configuration page, then double-click InSite ExC Agent Configuration.
NOTE:
Remote access is ONLY possible if the Service Platform (InSite) is properly configured (either by the
user or a GE technician at site). Operation see
Section 3-12-7 "Insite Exc Configuration" on page 3-86
7-6-5-2
Asset Performance Management (APM)
APM enables upload of system usage data for asset performance management purposes. The data
contain protected health information.
When connected (APM is active), the Voluson S&P-Series system has the capability to transmit certain
data sets to the General Electric Back Office. These data sets pertain to system utilization and are
accessible via proprietary tools only. Upon request, General Electric can provide a complete listing of
data sets gathered.
Data sets will be transmitted during the warranty period and when the Voluson S&P-Series system is
under a maintenance contract with General Electric. The requirement for data transfer is a signed
service agreement with valid data transfer clauses
The Field Engineer has the ability to stop the data transmission. The data transfer needs to be manually
enabled by the Field Engineer if there is an active contractual agreement for the data transfer.
APM structure datasets (*.xml) are stored in the D:/InSiteUploadFolder.
Figure 7-38 CSD - Configuration - Asset Performance Management