Section 4-5 - Backup and Restore Database, Preset Configurations and Images
4-5-2-3
Load only parts of the “Complete Backup”
NOTE:
Following procedure should be used, if the user has saved own 2D/3D/4D Programs or AutoText
entries, but new presets have to be added to match the installed Application Software version (e.g., 2D/
3D/4D Programs for new probe, etc.).
2.) Click the + sign (next to “Complete Backup”) to open the content tree.
3.) For example: Click the + sign and copy the desired content by clicking the >> button; and so on
NOTE:
To return selected items from the “Load Data” field to “Backup Data” field select the
<<
button.
4.) Confirm selection with the LOAD button.
Settings will be loaded and the US Application Software restarts.
Figure 4-15 Load image settings for probe (e.g., RAB4-8-D)
copy desired
content into
right field
[+] sign
>> button
[Load] button