GE H
EALTHCARE
DRAFT
V
OLUSON
E8 / V
OLUSON
E6
D
IRECTION
KTD102576, R
EVISION
7
DRAFT (A
UGUST
23, 2012)
S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-19
Section 8-10
Replacement of the Probe Holder (Kit)
8-10-1
Manpower
One person, 1 minute
8-10-2
Tools
None
8-10-3
Probe Holder (Kit)- Removal Procedure
1.) Pull out the elastic probe holder to be replaced.
8-10-4
Probe Holder (Kit) - Installation Procedure
1.) Insert the new probe holder from the kit.
Figure 8-20 pull out the probe holder
pull out the probe holder