Section 8-7 - Replacement or Activation of Options
Section 8-7 Replacement or Activation of Options
8-7-1
Operation for activating Options
1.) Press the
UTILITIES
key on the control panel.
2.) In the “Utilities” menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION and then click the
OPTIONS
tab.
Voluson E6 BT12 systems
Voluson E8 BT12 systems
Advanced 4D
Advanced 4D
DICOM
DICOM
VOCAL II
VOCAL II
Advanced VCI (Volume Contrast Imaging)
Advanced VCI (Volume Contrast Imaging)
Elastography
Elastography
Advanced SRI (Speckle Reduction Imaging)
Advanced STIC (Spatio-Temporal Image Correlation)
STIC (Spatio-Temporal Image Correlation)
SonoVCAD Labor
SonoVCAD Labor
Contrast (Coded Contrast Imaging)
Contrast (Coded Contrast Imaging)
SonoAVC (Sono Automated Volume Count)
SonoVCAD Heart
Expert (= Upgrade Voluson E8 -> Voluson E8 Expert)
SonoAVC (Sono Automated Volume Count)
Anatomical M-Mode
Anatomical M-Mode
Voluson E6 BT13 systems
Voluson E8 BT13 systems
Advanced 4D
Advanced 4D
DICOM
DICOM
SonoIT
SonoIT
VOCAL II
VOCAL II
Advanced VCI (Volume Contrast Imaging)
Advanced VCI (Volume Contrast Imaging)
Elastography
Elastography
Elastography Analysis
Elastography Analysis
STIC (Spatio-Temporal Image Correlation)
V-SRI (Volume Speckle Reduction Imaging)
SonoVCAD Labor
Advanced STIC (Spatio-Temporal Image Correlation)
Contrast (Coded Contrast Imaging)
SonoVCAD Labor
SonoVCAD Heart
Contrast (Coded Contrast Imaging)
SonoAVC (Sono Automated Volume Count)
SonoAVC (Sono Automated Volume Count)
Anatomical M-Mode
Expert (= Upgrade Voluson E8 -> Voluson E8 Expert)
SonoBiometry
Anatomical M-Mode
SonoBiometry