Chapter 4 - General Procedures and Functional Checks
4-19
P R E L I M I N A R Y
4-3-3-3
Adjust the 2D Mode Controls
The following controls can be adjusted to optimize the 2D Mode display:
•
Swipe to page 2 on the Touch panel and press either
Soft
or
Sharp
Auto Tissue setting.
-
Soft
: optimizes the radial and lateral uniformity and brightness of the tissue continuously in
real-time.
The mention “Soft” is displayed on the upper right corner of the image area
-
Sharp
: further enhances the image display by optimizing the gray scale curve.
The mention “Sharp” is displayed on the upper right corner of the image area
The Auto Tissue setting (Soft or Sharp) can be turned on/off by pressing
Auto
on the Control panel.
The last used setting is then applied.
The Auto Tissue settings are only available in live scanning and cannot be turned off when the
image is stored.
•
If available, press
Virtual Apex
(probe dependent) to improve near field imaging, allowing
increased visibility up to the width of the full probe aperture close to the surface.
•
Use the
Gain
and
TGC
controls to optimize the overall image.
Gain increases or decreases the amount of echo information displayed. TGC compensates for
depth-related attenuation in the image.
•
Use the
Depth
control to adjust the range to be imaged.
•
Use the
Frequency
control (move to higher frequencies) or the
Frame rate
control (move to lower
frame rate) to increase resolution in image.
•
Use the
Frequency
control (move to lower frequency) to increase penetration.
•
Use the
Reject
control to reduce noise in the image.
•
Use the
DDP
control to optimize imaging in the blood flow regions and make a cleaner, less noisy
image.
•
Use
UD Clarity
(Cardiac) or
UD Speckle reduce
(non-cardiac) to reduce image speckle. Extra care
must be taken to select the optimal Speckle reduction level, as too much filtering of speckle can
mask or obscure desired image detail.
•
Adjust
Octave
to toggle between fundamental and Harmonic mode.
•
Press
Color maps
and select a grey map from the menu on screen.
•
If using a 4D probe:
-
Adjust the
Quick Rotate
control on the Touch panel or press
Angle
on the Control panel to
rotate the scan plane to predefined angles.
-
Adjust the
Rotate
rotary of the Touch panel to fine tune the angle adjustment.
A scan plane indicator is displayed showing the angle position of the scan plane.
WARNING
WARNING
Always use the minimum power required to obtain acceptable
images in accordance with applicable guidelines and policies.
Содержание Vivid S60N
Страница 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Страница 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Страница 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Страница 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Страница 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Страница 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Страница 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Страница 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Страница 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Страница 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Страница 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Страница 477: ......