Chapter 10 - Care and Maintenance
10-3
Section 10-2
Why Perform Maintenance Procedures?
10-2-1
Keeping Records
It is good business practice that ultrasound facilities maintain records of quality checks and corrective
maintenance. The records provide the customer with documented confirmation that the Vivid q N
portable ultrasound scanner is maintained on a periodic basis.
Copies of the records should be kept in the same room as the
Vivid q N ultrasound scanner, or nearby.
10-2-2
Quality Assurance
In order to gain accreditation from organizations such as the
American College of Radiology
(
USA
), it
is the customer’s responsibility to have a Quality Assurance program in place for each scanner. The
program must be directed by a medical physicist, the supervising radiologist/physician or appropriate
designee.
Routine Quality Control testing of the system must be conducted regularly. The same tests are
performed regularly during each period, so that changes can be monitored over time and effective
corrective action taken, if required.
Testing results, corrective action, and the effects of corrective action, must be documented and
maintained on site.
Your GE Service Representative can help you with establishing, performing and maintaining records for
a Quality Assurance program. Please contact us for details of your nearest
GE Service Representative, who will provide a quotation for this service, as required.
Содержание Vivid q N
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type...
Страница 697: ......
Страница 698: ...2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...