Section 7-2 - Diagnostics
•
RFI Audio Test
This tests the audio frequency for Doppler sound (coming from the RFI board to the Back End).
During this test - approximately midway on the progress bar - a loud, high-pitched sound will be
heard coming from the speakers.
For procedure details, see the
RFI Audio Test section, page 7-43
.
NOTE:
To conduct a complete audio test of the system, it is also necessary to perform an audio test
which is part of the Back End - for details, refer to the
7-2-4-2-3
TR32 Tests
•
Beamformer Tests
-
TR32 BF Digital Test
This is an internal test of the Beamformers. Each Beamformer generates an internal test
pattern (first on the TR32A board and then on the TR32B board). This test pattern is
generated through MLA0 and MLA1 to the RFI.
Therefore, in the first phase of the test, the Beamformer on the TR32A board generates a test
pattern signal (while TR32B generates pattern
zero
).
In the second phase of the test, the Beamformer on the TR32A board generates a test pattern
zero
- while the BF on the TR32B board generates a test pattern signal.
Both the test patterns are summed and focused on the BF 32A and generated to the RFI
demodulator.
For test procedure instructions, see the
TR32 BF Digital Test section, page 7-46
-
System BF Digital Test
In this test, both Beamformers are generating internal test patterns. The patterns from the
BF of the TR 32B board are summed and focused in the BF of the TR32A board. From here,
the pattern is generated through MLA0 and MLA1 to the RFI.
For test procedure instructions, see the
System BF Digital Test section, page 7-49
-
ADC Test
The ADC Test is used to check that 1st harmonic, 2nd harmonic, 3rd harmonic, and
background noise are within pre-defined limits.
In this test, a signal is generated from the RFI into the TR switches. For all 64 channels of the
system, each signal goes through the following: pre-amplification, filtering, analog-to-digital
conversion, and focusing. The test signal is at a frequency of 2 MHz.
For test procedure instructions, see the
•
Receive (RX) Tests
-
RX Test
The RX Test is performed with no probe connected. This is used to check that the signal
values at the input to the TR32 board are as expected.
When performing the RX Test, test signals are generated from the RFI, and the results are
expected to be of fairly high amplitudes with no distortion of the frequency.
For test procedure instructions, see the
.
Содержание Vivid q N
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type...
Страница 697: ......
Страница 698: ...2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...