Section 5-3 - Front End Unit
5-3-5
Probe and MUX (P&M) Board
The Probe and MUX (multiplexer) Board enables acquisition and processing of signals from and to
probes connected to the front panel. The P&M enables up to 64 receive/transmit channels, and
performs the following functions:
•
Receiving the channels and connecting them to the TR32 Boards
•
Probe connectors
•
High voltage MUX that selects 64 out of 128 inputs, routed to the TR32 boards
•
Pencil's probe Pulser and receive pre-amplifier
•
Analog Doppler processor: a signal path for CW mode containing analog Beam Former, Band Pass
filter, I/Q base band converter, high pass filter.
Figure 5-14 Probe and MUX Board - Block Diagram
Содержание Vivid q N
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type...
Страница 697: ......
Страница 698: ...2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com...