Section 7-6 - Troubleshooting
7-6-1-8
Scan Procedure
Figure 7-14 Unable to Scan
Unable to scan on any of the Ports!
Check the probe connection.
YES
NO
Is probe connected in right
way?
Whether the system can
display the probe name?
Whether the probe
connected well?
NO
YES
Is the system in
Freeze status?
YES
UnFreeze the system.
NO
Go to the next page.
NO
Reconnect the
probe.
YES
1) Check Connector
Board,TX,RX and TMST.
2)Check TMST to RX Flex
cable
Whether the system
can be unfreezes?
NO
1) Check KBD.
2) Check TMST.
3) Check the wire connection
YES