Chapter 8 - Replacement procedures
8 - 87
8-6-4-4
Install the LCD Arm assembly
Follow these steps to install the LCD Arm Assembly:
1.) Carefully install the LCD Arm assembly into position, first feeding the LCD Arm cables down through
the console opening.
2.) Push the LCD Mount Lock Handle into locked position.
Figure 8-94 LCD Arm installed onto the UI Frame Upper (Arm for 17” screen illustrated)
Figure 8-95 LCD Mount Lock Handle
DO NOT ADJUST THIS NUT!