GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics / troubleshooting
7 - 11
7-6-2
Collect a Trouble Image with Logs
If the system should malfunction, press the
Alt+D
keys simultaneously. This will collect a screen capture
of the monitor, system presets and several log files in a date and time stamped “.zip” file.
NOTE:
This function may also be used to make a Print Screen.
The
Alt+D
function is available at all times.
When
Alt+D
is pressed, a menu box appears that allows for;
•
a place to enter a description of the issue
•
a check box to indicate a System lockup
•
a choice to Export to a pre-formatted CD-R/DVD-R or save to the Export directory D: drive (for
remote viewing through InSite).
NOTE:
You
MUST
select one of the available devices as the destination device if it is to be different
than the default Export directory on the hard drive.
The screen capture is a bitmap which eliminates the possibility of artifacts from compression.
7-6-2-1
Advanced log options
•
Extensive Log
enables the creation of a log file containing additional information for the selected
functionality.
•
Options
enables creation of a log file based on a selected bookmark or for a user configurable time
frame. Different type of information can be selected to be part of the log file.
Figure 7-1 System problem report (ALT+D dialog box)
TYPE DESCRIPTION OF ISSUE HERE
SELECT IF YOU’VE HAD A SYSTEM
LOCKUP (AFTER RESTART)
SELECT WHERE TO STORE THE
REPORT
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......