Section 4-2 - General procedures
4-2-3-5
Manually releasing the XY Lock
Even if the power is OFF, it is possible to manually release the console’s XY-mechanism (frog leg).
The release point is located on the rear of the XY Mechanism (Frog Leg).
1.) Insert a small tipped screwdriver or eq. into the release point and push until a “click” is heard. Some
force may be required if the lock adjustment is tight adjusted.
2.) Move the console to the wanted position.
Figure 4-10 XY mechanism release
Figure 4-11 Manually releasing the XY Lock
XY RELEASE