Chapter 8 - Replacement Procedures
8 - 79
Restore Standard Printer IP and Port Settings - Software Version v2.x (cont’d)
9.) Verify
Printer Name or IP Address
as recorded in
.
10.)Select
Raw
.
11.)Verify
Port Number
as recorded in
12.)Select
ok
.
13.)Close the rest of the Windows by selecting
ok
.
8-9-16-1
DC Offset Calibration
1.) Disconnect all probes.
2.) From the upper section of the window, select
Test
.
3.) Then select
Go to System Test.
4.) From the Module selection block, select
DC Offs. Cal
.
5.) Reboot the System when the DC Offset Calibration has been completed.
8-9-16-2
Verify System Software Version
1.) Press
CONFIG
(
F2
) and log on as
adm
, see
4-2-4 "Log On to the System as ‘ADM’" on page 4-12
2.) Select
About
tab.
3.) Verify that the SW versions are the same as what were recorded in
and hardware versions" on page 8-69
.
8-9-17
Verification - Functional Checks
•
Please perform the tests specified in
Section 8-20 "Verification - Functional Checks" on page 8-105
Figure 8-77 Configure Port
Raw
Port Number
ok
Printer Name or IP Address
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......