Section 7-8 - Acquisition Diagnostics
7-8-3-8
Receiver Board (RX) Test
7-8-3-9
Relay Board Test
NOTE:
A Test Attenuator, Part Number AA200215 is needed for this test.
Table 7-9
Receiver Board (RX) Test
Test (Key) Sub-tests
Description
RX
Receiver Board (RX)
Test All
Activates the following tests, in sequence, see description for each sub test;
- RECV0
- RECV1
RECV.0
Max Gain test for RX using FEC signal generator.
RECV.1
Reduced Gain test for RX using FEC signal generator.
Table 7-10 Relay Board (RLB) Test
Test (Key) Sub-tests
Description
RLB
Relay Board (RLB)
Test All
Activates the following tests, in sequence (see the descriptions for the tests below);
1. Conn. 1
2. Conn. 2
3. Conn. 3
Conn. 1
Max Gain test for RX using the signal generator on the Front-End Controller (FEC) and a Test
Attenuator, Part Number AA200215, in left (active) probe connector.
Conn. 2
Max Gain test for RX using the signal generator on the Front-End Controller (FEC) and a Test
Attenuator, Part Number AA200215, in mid (active) probe connector.
Conn. 3
Max Gain test for RX using the signal generator on the Front-End Controller (FEC) and a Test
Attenuator, Part Number AA200215, in right probe connector.
Содержание Vivid 7
Страница 1: ...GE Healthcare Operating Documentation Vivid 7 Service Manual Part Number FC091194 Revision 11...
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL vii JA ZH CN KO...
Страница 37: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxv Site Log 10 27 Index Index 1...
Страница 38: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL xxxvi...
Страница 856: ...GE HEALTHCARE DIRECTION FC091194 REVISION 11 VIVID 7 SERVICE MANUAL Index 8...
Страница 857: ......