Set the Remote Archive’s Network Information - Software v2.x
In the Remote Archive Setup area of the screen (see example in
.), enter the;
1.) Remote Archive IP address. (Default IP Address from factory: 10.0.0.4).
2.) Remote Archive Name. (Default Remote Archive Name from factory: ECHOPAC7-000001).
.
3-11-7
Save the New Settings - Software v2.x
1.) Press
SAVE SETTINGS
to save the new settings.
The new settings are saved to a common settings file. After a restart, the settings are also included in
other screens.
2.) Restart Vivid 7 to make the changes take place.
Figure 3-120 TCP/IP Set-up for Remote Archive (EchoPAC PC)
Remote Archive
IP address
Remote Archive
Name
Remote Archive
Setup area
Save settings