DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
Chapter 8 - Replacement Procedures
8-13
P R E L I M I N A R Y
8-2-7
Fan Replacement Procedure
8-2-7-1
Tools
Phillips screwdriver of appropriate size
8-2-7-2
Time Required
5 min
8-2-7-3
Preparations
1.) Shut down the Venue Fit™ ultrasound unit, as described in
.
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Fit™ system from cart as described in
Mounting / Dismounting System on Cart
and place it on a previously prepared surface with the display facing down.
8-2-7-4
Fan Removal Procedure
1) Remove the Rear cover.
2) Remove the Speakers.
3.) Remove M-PSB.
•
•
•
Содержание Venue Fit
Страница 299: ......