Preparing the System for Use
2-4
Venue 50
–
Basic User Manual
Direction 5446729-100 English
Rev. 7
Environmental Requirements
The system should be operated, stored, or transported within
the parameters outlined below. Either its operational
environment must be constantly maintained or the unit must be
turned off.
Acclimation Time
After being transported, the unit requires one hour for each 2.5
degree increment when its temperature is below 10°C or above
40°C.
Table 2-1: System Environmental Requirements
Operational
Storage
Transport
Temperature
3 - 40°C
37 - 104°F
-5 - 50°C
23 - 122°F
-5 - 50°C
23 - 122°F
Humidity
30 - 80% non-condensing
10 - 90% non-condensing
10 - 90% non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
Table 2-2: System Acclimation Time Chart
°C
60
55
50
45
40
35
30
25
20
15
10
°F
140
131
122
113
104
95
86
77
68
59
50
hours
8
6
4
2
0
0
0
0
0
0
0
°C
5
0
-5
-10
-15
-20
-25
-30
-35
-40
°F
41
32
23
14
5
-4
-13
-22
-31
-40
hours
2
4
6
8
10
12
14
16
18
20
Содержание Venue 50
Страница 4: ...i 2 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7 This page intentionally left blank...
Страница 8: ...i 6 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 14: ...i 12 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 56: ...Introduction Safety 1 42 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 236: ...Customizing Your System 4 34 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 280: ...Probes and Biopsy 5 44 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 327: ......