Calibration
5-3
March 2004
Trimming 4-20 mA via
the Keypad
1.
In the
User Program
, scroll to
CAL
with
4-20 Loop
highlighted.
Press [ENT].
2.
Scroll to
Mode
and press [ENT]. In the
Mode
window, scroll to
Test[Trim]
and press [ENT].
3.
Return to the
Loop
window, and scroll to
Percent.
4.
In the
Percent
window, use the arrow keys to set the percentage to
100%. Press [ENT]. Record the value shown on the current meter.
5.
Now use the arrow keys to set the percentage to 0%. Press [ENT].
Record the value shown on the current meter.
6.
Return to the
Loop
window, and scroll to
Base Trim
. Press [ENT].
7.
Use the arrow keys to enter the base trim value, the value recorded
in step 5. (The loop current should now be the same as the current
meter. It should read 4.0 +/-0.01 mA.) Press [ENT]
8.
Repeat steps 3 and 4.
9.
Return to the
Loop
window, and scroll to
Span Trim
. Press [ENT].
10.
Use the arrow keys to enter the span trim value, the value recorded
in step 4. (The loop current should now be the same as the current
meter. It should read 20.0 +/-0.01 mA.) Press [ENT].
11.
Return to the
Loop
window, and scroll to
Mode
. Press [ENT].
Scroll to
Normal,
and press [ENT].
Содержание UTX878 Ultrasonic
Страница 1: ...March 2004 Process Control Instruments UTX878 Ultrasonic Flow Transmitter User s Manual 910 253A ...
Страница 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Страница 8: ...Chapter 1 ...
Страница 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Страница 21: ...March 2004 Installation 1 12 Figure 1 6 Clamping Fixture Outline and Installation Dwg 570 076 rev 1 ...
Страница 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Страница 23: ...Chapter 2 ...
Страница 45: ...Chapter 3 ...
Страница 52: ...Chapter 4 ...
Страница 60: ...Chapter 5 ...
Страница 65: ...Chapter 6 ...
Страница 75: ...Chapter 7 ...
Страница 80: ...Appendix A ...
Страница 85: ...Appendix B ...
Страница 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Страница 88: ...Appendix C ...
Страница 89: ...Data Records Site Data C 1 ...
Страница 93: ...Appendix D ...
Страница 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...
Страница 96: ...March 2004 D 2 Service Record Table D 1 Service Record cont Date Description of Service Performed By ...