Chapter 5. Advanced Programming
58
TMO2D Digital Display User’s Manual
5.7 Response Parameters? (cont.)
The next setting, Tau (down), is the tau value for calculations while the measured gas concentration is decreasing.
Increasing the Tau (down) reading increases the transmitter’s response speed.
5.8 LabCal Outputs?
The
LabCal Outputs?
option enables you to calibrate (or trim) the analog outputs. After you select the output (A or B)
to calibrate, the output is driven to 20 mA. You can input the actual output value (as shown by a multimeter or
connected control device) and adjust the output. The output is then driven to 4 mA, and you can make a similar
adjustment.
FACTORY SETUP MENU
Tau (down) [XX.XX]:
Use the numeric keys to enter the desired value. Press [YES] to enter the number,
and [YES] again to confirm the entry.
FACTORY SETUP MENU
Thresh (%) [XX.XX]:
Use the numeric keys to enter the desired value. Press [YES] to enter the number,
and [YES] again to confirm the entry.
FACTORY SETUP MENU
Smooth (%) [XX.XX]:
Use the numeric keys to enter the desired value. Press [YES] to enter the number,
and [YES] again to confirm the entry.
FACTORY SETUP MENU
Window (%) [XX.XX]:
Use the numeric keys to enter the desired value. Press [YES] to enter the number,
and [YES] again to confirm the entry.
FACTORY SETUP MENU
Response Parameters?
Press [NO] to proceed to the LabCal Outputs? option.
FACTORY SETUP MENU
LabCal Outputs?
Press [YES] to calibrate the analog outputs.
Output to LabCal:
[A] B done
Use the [NO] or arrow keys to scroll to the desired output. Press [YES] to confirm
the entry.
Output A 20 mA Output
Enter mA [20.00]:
Use the numeric keys to enter actual output value (as read by a multimeter or control
device). Press [YES] to enter the number, and [YES] again to confirm the entry.
Output A 4 mA Output
Enter mA [4.00]:
Use the numeric keys to enter actual output value (as read by a multimeter or control
device). Press [YES] to enter the number, and [YES] again to confirm the entry.
Output to LabCal:
A [B] done
Use the [NO] or arrow keys to scroll to another output to repeat the calibration
procedure. Press [YES] to confirm the entry. If you have completed calibration,
scroll to “Done” and press [YES].
FACTORY SETUP MENU
LabCal Outputs?
Press [NO] to proceed to the Manual Offset? option.
Содержание TMO2D
Страница 1: ...GE Measurement Control Solutions Gas Moisture TMO2D Digital Display User s Manual ...
Страница 3: ...GE Measurement Control Solutions TMO2D Digital Display User s Manual 910 084 Rev D August 2011 ...
Страница 4: ...ii ...
Страница 10: ...Chapter 1 Features and Capabilities 2 TMO2D Digital Display User s Manual ...
Страница 24: ...Chapter 2 Installation 16 TMO2D Digital Display User s Manual ...
Страница 26: ...Chapter 2 Installation TMO2D Digital Display User s Manual 18 ...
Страница 72: ...Chapter 6 Specifications 64 TMO2D Digital Display User s Manual ...
Страница 74: ...Appendix A Outline and Installation Drawings 66 TMO2D Digital Display User s Manual ...
Страница 86: ...Appendix B Menu Maps 78 TMO2D Digital Display User s Manual ...
Страница 92: ...Appendix B Menu Maps TMO2D Digital Display User s Manual 84 ...
Страница 96: ...Index 88 TMO2D Digital Display User s Manual ...
Страница 98: ...Warranty 90 TMO2D Digital Display User s Manual ...
Страница 100: ......
Страница 101: ......